TruSystem Surgical Table (TS7000)  Leg Section One Part Light HU Use of the Leg Section One Part Light HU is permitted only with the TRUMPF TruSystem 7000 operating table. The operating table TruSystem 7000, in conjunction with operating table components is intended for the use of patient positioning during surgery, ranging from anesthesia induction to actual surgery to recovery from anesthesia. The Leg Section One Part Light HU is attached to coupling point L on the operating table and is used to position the patients legs during pre-op and post-op surgery.

Published November 27, 2013 · Recall initiated: October 28, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The coupler point (hood) of the Leg Section One Part Light may become detached from the leg section frame.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Trumpf Medical Systems, Inc.
Distributor
US Distribution including the states of CA, IL, NJ, NY, NC, TX, WA and SC.
Lot
Item no. 1739993, All serial numbers.
Risk classification
Class II
Area (FDA)
device
Country
United States