Back to search
FDAFDA · Food and drugs·Recall number Z-0342-2022

Puritan Bennett 980 Series Ventilator, 980U1ENASAA

Published December 15, 2021 · Recall initiated: November 4, 2021

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Manufacturing assembly error where a capacitor within the ventilator was assembled incorrectly, which may cause the device to become inoperable during use.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Covidien Llc
Distributor
Worldwide distribution - US Nationwide.
Lot
GTIN/UDI - 10884521201385 Serial Numbers: 35B1700252 35B1700365 35B1700399 35B1700407 35B1700489 35B1700505 35B1700513 35B1700555 35B1700557 35B1700562 35B1700563 35B1700565 35B1700566 35B1700605 35B1700612 35B1700634 35B1700651 35B1700662 35B1700663
Risk classification
Class I
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card