Back to search
FDAFDA · Food and drugs·Recall number Z-0375-2014

Stryker 1) XIA Torque Wrench 2) XIA 3 Torque Wrench 3) XIA Elegance Short Torque Wrench Manufactured by: Stryker Spine SAS Z.I. Marticot -33610 -CESTAS -FRANCE +33 (0) 5.57.97.06.30 http://www.stryker.com Distributed in the USA by: Stryker Spine 2 Pearl Court Allendale, NJ 07401-1677 USA +1-201-760-8000 XIA 3 Spinal System- Intended for use in the noncervical spine. When used as an anterior/anterolateral and posterior, non cervical pedicle and on-pedicled fixation system, the XIA 3 Spinal system is intended to provide additional support during fusion autograft or allograft in skeletally mature patients in the treatment of the following acute and chronic instabilities or deformities; degenerate disc disease (DDD) ; spondylolisthesis; trauma spinal stenosis; curvatures; tumor; pseudoarthrosis and failed previous fusion. XIA - Intended for anterior/anterolateral and posterior, non cervical pedicle and non-pedicle fixation of the spine to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion for the following indications: degenerative disc disease; spondylolisthesis, trauma, spinal stenosis, curvatures, tumor, pseudoarthrosis and failed previous fusion.

Published November 27, 2013 · Recall initiated: September 24, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Multiple complaints have reported that the hex tip of the torque wrench fracturing during final tightening.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Stryker Spine
Distributor
Worldwide Distribution - USA (nationwide) and Internationally to Argentina, Australia, Canada, Chile, France, Canada, Germany, Hong Kong, Ireland, Italy, Japan, Korea, Netherlands, New Zealand, Spain, Iraq, Egypt, and Turkey.
Lot
Medical Device Listing Number E207315 1) XIA Torque Wrench Catalog # 03807028 Lot # 127510, 092793, 097089, 11E038, 11E039, 11E040, 11E041, 123746, 127179, &12D012 2) XIA 3 Torque Wrench Catalog # 48237028 Lot # 11E042, 11E044, 11E047, 098525, 118823, 11E043, 11E045, 11E046, 11E048, 127647, R127647, 12A646, & R12A646 3) XIA Elegance Short Torque Wrench Catalog # 482397028 Lot # 11A957, 11E035, 11E036, R11E036, & 121098 4) Mantis Redux Torque Wrench Reference Number 48287028 Lot # 098401, 125708, 125709, 127051 & 12A641
Risk classification
Class II
Area (FDA)
device
Country
United States