FDAFDA · Food and drugs·Recall number Z-0379-2018
Metrotom 800 (130kV CT scanner) Industrial CT scanner for measuring and inspecting complete components made of plastic or light metal.
Published February 7, 2018 · Recall initiated: November 14, 2017
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to a software error, it was observed that when the user closes the access door, the system will resume its scan cycle automatically, instead of the user initiating x-ray generation from the control panel.
What to do
FDA Mandated · Letter
Recall details
- Manufacturer
- Carl Zeiss Metrology Inc
- Distributor
- Worldwide Distribution - US, including the states of : WI, NC, GA, IL, AZ, FL, CA, NJ, and territory of PR; and the country of Canada.
- Lot
- Serial Numbers: 152806 157499 157683 181193 181364 182887 186654 186767 186797 192099 192182 192351 197342
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States