Metrotom 800 (130kV CT scanner) Industrial CT scanner for measuring and inspecting complete components made of plastic or light metal.
Published February 7, 2018 · Recall initiated: November 14, 2017
Summary
FDA records medical device Class II recall Z-0379-2018, published February 7, 2018 and initiated November 14, 2017. Recalling firm: Carl Zeiss Metrology Inc; Status: Ongoing; Reported quantity: 13 US; Lot: Serial Numbers: 152806 157499 157683 181193 181364 182887 186654 186767 186797 192099 192182 192351 197342. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Due to a software error, it was observed that when the user closes the access door, the system will resume its scan cycle automatically, instead of the user initiating x-ray generation from the control panel.
What to do
FDA Mandated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Carl Zeiss Metrology Inc
- Distributor
- Worldwide Distribution - US, including the states of : WI, NC, GA, IL, AZ, FL, CA, NJ, and territory of PR; and the country of Canada.
- Lot
- Serial Numbers: 152806 157499 157683 181193 181364 182887 186654 186767 186797 192099 192182 192351 197342
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAWegmans Ready To Cook Shrimp Scampi, 6 oz. UPC: 2-16151-00000February 7, 2018
- FDAAffixus Hip Fracture Nail, 125 DEG 11MM X 180MMFebruary 7, 2018
- FDAWegmans Ready To Cook Lobster & Shrimp Steam Pot, Sold by the lb. UPC: 2-16326-00000February 7, 2018
- FDAAffixus Hip Fracture Nail, 130 DEG 9MM X 180MMFebruary 7, 2018
- FDASun Noodle brand Tonkotsu Ramen, Packaged in plastic clear clamshell, Banded with Tonkotsu label,February 7, 2018
- FDAAffixus Hip Fracture Nail, LH 125 DEG 11MM X 360MMFebruary 7, 2018
Questions about this recall
Was Metrotom 800 (130kV CT scanner) Industrial CT scanner for measuring and inspecting complete components made of plastic or light metal. recalled?
Yes. FDA published recall Z-0379-2018 on February 7, 2018. Reported hazard: Due to a software error, it was observed that when the user closes the access door, the system will resume its scan cycle automatically, instead of the user initiating x-ray generation from the control panel.
What should I do if I have this product?
FDA Mandated · Letter