SPECIALTY AUDIBLE TORQUE WRENCH, XIA IS2217XLP MFG BY: STRYKER SPINE 2 PEARL COURT ALLENDALE NJ 07401-1677 The Xia 3/Mantis Redux/Specialty Audible Torque Wrenches are used to final tighten the blocker of the Xia 3/ Mantis Redux Polyaxial Screws, Hooks, Monoaxial Screws, Rod to Rod Connectors and Off-set connectors. The Torque Wrench is a t-handle deflection beam torque wrench with a 5 mm hex tip. The instrument allows the controlled application of 12 Nm torque via a visual indicator on the bell shaped cylinder just below the handle. The Specialty Audible Torque Wrench has an extra feature where it creates and audible "click" when the desired final tightening torque is reached.
Published November 28, 2012 · Recall initiated: August 9, 2012
Summary
FDA records medical device Class II recall Z-0382-2013, published November 28, 2012 and initiated August 9, 2012. Recalling firm: Stryker Spine; Status: Terminated; Reported quantity: 28 units; Lot: Class I device, 510(K) exempt Catalog Number: IS2217XLP Lot Number: 2576, 2756. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Multiple complaints have reported the hex tip of the torque wrench fracturing during final tightening.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Stryker Spine
- Distributor
- Worldwide Distribution - USA (nationwide) and internationally to Australia, Canada, France, Germany, Netherlands, Japan, Spain, Denmark, Switzerland, and the United Kingdom.
- Lot
- Class I device, 510(K) exempt Catalog Number: IS2217XLP Lot Number: 2576, 2756
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAVS Anchor-C Cervical Cage System Instructions For Use (IFU)- eIFU (revision 5)-Indicated for anteriorOctober 2, 2024
- FDAStryker Radius 5.5 to 3.5 Rod to Red Connector, REF 48664530. Intended to provide additional support duringNovember 18, 2015
- FDASpecialty Audible Torque Wrench Mfg by: Stryker Spine, Catalog number IS3002XLP, Non-sterile. Used to finalDecember 24, 2014
- FDASpecialty Aria Inserter, REF IS3021ARI, Non-Sterile, This instrument is part of the Aria product brand. It isSeptember 17, 2014
- FDAStryker MANTIS TORQUE WRENCH REF 48287028 Manufactured by: Stryker pine SAS Z.I MARTICOT-33610-CESTAS-FRANCENovember 28, 2012
- FDA"***SENTINEL***Cannulated Drill Bit***S8580***8 X 229mm (9")***CONMED LINVATEC 11311 CONCEPT BLVD LARGO, FLNovember 28, 2012
Questions about this recall
Was SPECIALTY AUDIBLE TORQUE WRENCH, XIA IS2217XLP MFG BY: STRYKER SPINE 2 PEARL COURT ALLENDALE NJ 07401-1677 The Xia 3/Mantis Redux/Specialty Audible Torque Wrenches are used to final tighten the blocker of the Xia 3/ Mantis Redux Polyaxial Screws, Hooks, Monoaxial Screws, Rod to Rod Connectors and Off-set connectors. The Torque Wrench is a t-handle deflection beam torque wrench with a 5 mm hex tip. The instrument allows the controlled application of 12 Nm torque via a visual indicator on the bell shaped cylinder just below the handle. The Specialty Audible Torque Wrench has an extra feature where it creates and audible "click" when the desired final tightening torque is reached. recalled?
Yes. FDA published recall Z-0382-2013 on November 28, 2012. Reported hazard: Multiple complaints have reported the hex tip of the torque wrench fracturing during final tightening.
What should I do if I have this product?
Voluntary: Firm initiated · Letter