SPECIALTY AUDIBLE TORQUE WRENCH, XIA IS2217XLP MFG BY: STRYKER SPINE 2 PEARL COURT ALLENDALE NJ 07401-1677 The Xia 3/Mantis Redux/Specialty Audible Torque Wrenches are used to final tighten the blocker of the Xia 3/ Mantis Redux Polyaxial Screws, Hooks, Monoaxial Screws, Rod to Rod Connectors and Off-set connectors. The Torque Wrench is a t-handle deflection beam torque wrench with a 5 mm hex tip. The instrument allows the controlled application of 12 Nm torque via a visual indicator on the bell shaped cylinder just below the handle. The Specialty Audible Torque Wrench has an extra feature where it creates and audible "click" when the desired final tightening torque is reached.

Published November 28, 2012 · Recall initiated: August 9, 2012

Summary

FDA records medical device Class II recall Z-0382-2013, published November 28, 2012 and initiated August 9, 2012. Recalling firm: Stryker Spine; Status: Terminated; Reported quantity: 28 units; Lot: Class I device, 510(K) exempt Catalog Number: IS2217XLP Lot Number: 2576, 2756. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Multiple complaints have reported the hex tip of the torque wrench fracturing during final tightening.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Stryker Spine
Distributor
Worldwide Distribution - USA (nationwide) and internationally to Australia, Canada, France, Germany, Netherlands, Japan, Spain, Denmark, Switzerland, and the United Kingdom.
Lot
Class I device, 510(K) exempt Catalog Number: IS2217XLP Lot Number: 2576, 2756
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was SPECIALTY AUDIBLE TORQUE WRENCH, XIA IS2217XLP MFG BY: STRYKER SPINE 2 PEARL COURT ALLENDALE NJ 07401-1677 The Xia 3/Mantis Redux/Specialty Audible Torque Wrenches are used to final tighten the blocker of the Xia 3/ Mantis Redux Polyaxial Screws, Hooks, Monoaxial Screws, Rod to Rod Connectors and Off-set connectors. The Torque Wrench is a t-handle deflection beam torque wrench with a 5 mm hex tip. The instrument allows the controlled application of 12 Nm torque via a visual indicator on the bell shaped cylinder just below the handle. The Specialty Audible Torque Wrench has an extra feature where it creates and audible "click" when the desired final tightening torque is reached. recalled?

Yes. FDA published recall Z-0382-2013 on November 28, 2012. Reported hazard: Multiple complaints have reported the hex tip of the torque wrench fracturing during final tightening.

What should I do if I have this product?

Voluntary: Firm initiated · Letter