MAS Urinalysis Control Level 1, Multi-analyte control (Assayed and Unassayed), Catalog/Part Numbers: 024222; Product is manufactured and distributed by Microgenics Corporation, a subsidiary of Thermo Fisher Scientific, Clinical Diagnostic Division, Fremont, CA MAS Urinalysis is intended for use in the clinical laboratory as a control for qualitative and semi-quantitative procedures used in routine urinalysis testing. Assay values are provided for specific systems listed. The user can compare observations with expected ranges as a means of assuring consistent performance of reagent and equipment. The product is suitable for use as a control material for physiochemical, chemical, and microscopic methods of routine urinalysis. Urinalysis may be used in conjunction with commercially available standardized systems for urine microscopic analysis.
Published December 4, 2013 · Recall initiated: October 29, 2013
Summary
FDA records medical device Class III recall Z-0383-2014, published December 4, 2013 and initiated October 29, 2013. Recalling firm: Thermofisher Scientive; Status: Terminated; Reported quantity: 914 units; Lot: Lot Number: UA1514011. Class III means the product is not likely to cause adverse health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Via one (1) consumer complaint and in-house testing has confirmed that the product does not meet the 30 day open vial stability specifications for Lot: UA1514011.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Thermofisher Scientive
- Distributor
- Worldwide Distribution - USA (nationwide) including Puerto Rico and internationally to Hong Kong.
- Lot
- Lot Number: UA1514011
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDATri-Nex Healing Abutment, 0.50 mm x 5 mm height, Catalog No. HA-L-50-5. The NSI Implant System is intended toDecember 4, 2013
- FDASynthes Matrix Threaded Persuader Deformity is a comprehensive thoracolumbar pedical screw system designed toDecember 4, 2013
- FDAREF 200068 Security (TM) Enclosed Carpal Tunnel System Blade ( Single) ABS /Stainless Steel Sterile Device,December 4, 2013
- FDAACCU-CHEK Aviva Combo Blood Glucose Meter (as a part of the ACCU-CHEK Aviva Combo System)\ The ACCU-CHEK ComboDecember 4, 2013
- FDA2.4 and 2.7 MM Locking Screw SLF-TPNG with T8 Stardrive Recess 12 MM Fixation of fractures, osteotomies,December 4, 2013
- FDASingle Analyte Urine DAU Control (5 m L Norbuprenorphine, Level II control For In Vitro Diagnostic Use Only.December 4, 2013
Questions about this recall
Was MAS Urinalysis Control Level 1, Multi-analyte control (Assayed and Unassayed), Catalog/Part Numbers: 024222; Product is manufactured and distributed by Microgenics Corporation, a subsidiary of Thermo Fisher Scientific, Clinical Diagnostic Division, Fremont, CA MAS Urinalysis is intended for use in the clinical laboratory as a control for qualitative and semi-quantitative procedures used in routine urinalysis testing. Assay values are provided for specific systems listed. The user can compare observations with expected ranges as a means of assuring consistent performance of reagent and equipment. The product is suitable for use as a control material for physiochemical, chemical, and microscopic methods of routine urinalysis. Urinalysis may be used in conjunction with commercially available standardized systems for urine microscopic analysis. recalled?
Yes. FDA published recall Z-0383-2014 on December 4, 2013. Reported hazard: Via one (1) consumer complaint and in-house testing has confirmed that the product does not meet the 30 day open vial stability specifications for Lot: UA1514011.
What should I do if I have this product?
Voluntary: Firm initiated · Letter