00780303601 MIS 12/14 Slotted Femoral Head Provisional -3.5X36 00780303602 MIS 12/14 Slotted Femoral Head Provisional +0X36 00780303603 MIS 12/14 Slotted Femoral Head Provisional +3.5X36 00780303604 MIS 12/14 Slotted Femoral Head Provisional +7X36 00780303605 MIS 12/14 Slotted Femoral Head Provisional +10.5X36 The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty.

Published November 28, 2012 · Recall initiated: October 10, 2010

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Surgical technique instructions for the Versys and Slotted Provisional Femoral Heads have been enhanced to inform users that the head may become difficult to visually locate if it becomes displaced deep or medial to the iliopsoas tendon and provide non-invasive imaging methods that can be used to locate the head. To reduce incidence of sub-optimal fit between provisional femoral heads and the mat

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Zimmer, Inc.
Distributor
Worldwide distribution: USA (nationwide) and countries of: Asia, Barbados, Canada, Europe, Mexico, Puerto Rico, San Juan, and South America.
Lot
All lots distributed from lot 69998600 (2/20/1996) to present
Risk classification
Class II
Area (FDA)
device