STERRAD 100NX, Product Code: 10104 The STERRAD Steri lization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.

Published December 4, 2013 · Recall initiated: November 11, 2013

Summary

FDA records medical device Class II recall Z-0402-2014, published December 4, 2013 and initiated November 11, 2013. Recalling firm: Advanced Sterilization Products; Status: Terminated; Reported quantity: Total 4,381 units; Lot: Please refer to consignee list. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Advanced Sterilization Products (ASP) is recalling the STERRAD NX and STERRAD 100NX sterilization systems because ASP has determined that the bronchoscopes may contain a polyurethane material in the suction channel location that is not listed in the current claims for interior channels (lumens) for processing in STERRAD systems.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Advanced Sterilization Products
Distributor
Nationwide Distribution
Lot
Please refer to consignee list
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was STERRAD 100NX, Product Code: 10104 The STERRAD Steri lization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices. recalled?

Yes. FDA published recall Z-0402-2014 on December 4, 2013. Reported hazard: Advanced Sterilization Products (ASP) is recalling the STERRAD NX and STERRAD 100NX sterilization systems because ASP has determined that the bronchoscopes may contain a polyurethane material in the suction channel location that is not listed in the current claims for interior channels (lumens) for processing in STERRAD systems.

What should I do if I have this product?

Voluntary: Firm initiated · Letter