qube Compact Monitor (Patient Monitor) 91390. The monitor is 26.2 cm x 31.5 cm x 13.2 cm and weighs 4.1 kg (10.3 inches x 12.4 inches x 5.2 inches - 9 lbs). It has a 26.2 cm (12.1 inches) LCD display with 1024 x 768 resolution. The Spacelabs Healthcare Qube Compact Monitor (91390), functioning as either bedside or central monitors; passively displays data generated by 2 Spacelabs Healthcare parameter modules, Flexports interfaces, and other SDLC based products in the form of waveform and numeric displays, trends and alarms. Key monitored parameters available on the Model 91390 when employing the Spacelabs Command Module, consist of ECU, respiration, invasive and noninvasive blood pressure, Sp02, temperature and cardiac output. Additional parameters and interfaces to other systems are also available depending on the parameter modules employed. The Qube is intended to alert the user to alarm conditions that are reported by Spacelabs Healthcare parameter modules and/or other physiologic monitors via Flexport interfaces. These devices determine a) when an alarm condition is violated; b) the alarm priority (i.e. high, medium or low); c) alarm limits; and d) when to initiate and terminate alarm notifications. The patient monitors are also capable of displaying alarm conditions on other monitors that are on the network through the Alarm Watch feature. The Qube may also function as a generic display or computer terminal. As a generic display or terminal, the patient monitors allow network-based applications to open windows and display information on other networked monitors. The Qube is also designed to communicate with a variety of external devices such as displays, network devices, serial devices, user input devices, audio systems, and local/remote recorders. The Qube is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a hospital environment.
Published November 28, 2012 · Recall initiated: October 19, 2012
Summary
FDA records medical device Class II recall Z-0403-2013, published November 28, 2012 and initiated October 19, 2012. Recalling firm: Spacelabs Healthcare, Llc; Status: Terminated; Reported quantity: 742 total devices distributed; 530 with the U.S. and 212 Internationally.; Lot: Serial numbers for devices distributed in the US: 1390-000102, 1390-000103, 1390-000104, 1390-000105, 1390-000106, 1390-000107, 1390-000108, 1390-000109, 1390-000110, 1390-000111, 1390-000112, 1390-000113, 1390-000114, 1390-000115, 1390-000116, 1390-000119, 1390-000120, 1390-000150, 1390-000151, 1390-000152, 1390-000153, 1390-000154, 1390-000155, 1390-000156, 1390-000157, 1390-000158, 1390-000159, 1390-000160, 1390-000161, 1390-000162, 1390-000163, 1390-000164, 1390-000165, 1390-000166, 1390-000167, 1390-000168, 1390-000169, 1390-000170, 1390-000171, 1390-000172, 1390-000173, 1390-000174, 1390-000175, 1390-000176, 1390-000177, 1390-000178, 1390-000179, 1390-000180, 1390-000181, 1390-000182, 1390-000183, 1390-000184, 1390-000185, 1390-000186, 1390-000187, 1390-000188, 1390-000189, 1390-000190, 1390-000191, 1390-000192, 1390-000193, 1390-000195, 1390-000196, 1390-000197, 1390-000198, 1390-000199, 1390-000200, 1390-000201, 1390-000202, 1390-000203, 1390-000204, 1390-000205, 1390-000206, 1390-000207, 1390-000208, 1390-000209, 1390-000210, 1390-000211, 1390-000212, 1390-000213, 1390-000214, 1390-000215, 1390-000216, 1390-000217, 1390-000218, 1390-000219, 1390-000220, 1390-000221, 1390-000222, 1390-000223, 1390-000224, 1390-000225, 1390-000226, 1390-000227, 1390-000228, 1390-000229, 1390-000230, 1390-000231, 1390-000232, 1390-000233, 1390-000234, 1390-000235, 1390-000236, 1390-000237, 1390-000238, 1390-000239, 1390-000240, 1390-000241, 1390-000242, 1390-000243, 1390-000244, 1390-000245, 1390-000246, 1390-000247, 1390-000248, 1390-000249, 1390-000250, 1390-000251, 1390-000252, 1390-000253, 1390-000254, 1390-000255, 1390-000256, 1390-000257, 1390-000258, 1390-000259, 1390-000260, 1390-000261, 1390-000262, 1390-000263, 1390-000264, 1390-000266, 1390-000267, 1390-000268, 1390-000269, 1390-000270, 1390-000271, 1390-000272, 1390-000273, 1390-000274, 1390-000275, 1390-000276, 1390-000277, 1390-000278, 1390-000279, 1390-000280, 1390-000281, 1390-000282, 1390-000283, 1390-000284, 1390-000285, 1390-000286, 1390-000287, 1390-000288, 1390-000289, 1390-000296, 1390-000297, 1390-000298, 1390-000299, 1390-000300, 1390-000301, 1390-000302, 1390-000304, 1390-000305, 1390-000307, 1390-000308, 1390-000310, 1390-000311, 1390-000312, 1390-000314, 1390-000315, 1390-000316, 1390-000317, 1390-000318, 1390-000319, 1390-000321, 1390-000322, 1390-000323, 1390-000325, 1390-000341, 1390-000342, 1390-000343, 1390-000344, 1390-000345, 1390-000346, 1390-000347, 1390-000348, 1390-000349, 1390-000350, 1390-000351, 1390-000352, 1390-000353, 1390-000354, 1390-000355, 1390-000356, 1390-000357, 1390-000358, 1390-000359, 1390-000360, 1390-000361, 1390-000362, 1390-000363, 1390-000364, 1390-000365, 1390-000366, 1390-000367, 1390-000368, 1390-000369, 1390-000370, 1390-000371, 1390-000372, 1390-000373, 1390-000374, 1390-000375, 1390-000376, 1390-000377, 1390-000378, 1390-000379, 1390-000380, 1390-000381, 1390-000382, 1390-000383, 1390-000384, 1390-000385, 1390-000386, 1390-000387, 1390-000388, 1390-000389, 1390-000390, 1390-000391, 1390-000392, 1390-000393, 1390-000394, 1390-000420, 1390-000421, 1390-000422, 1390-000423, 1390-000424, 1390-000425, 1390-000426, 1390-000427, 1390-000428, 1390-000429, 1390-000430, 1390-000431, 1390-000432, 1390-000433, 1390-000434, 1390-000435, 1390-000436, 1390-000437, 1390-000438, 1390-000439, 1390-000440, 1390-000441, 1390-000442, 1390-000443, 1390-000444, 1390-000445, 1390-000446, 1390-000447, 1390-000448, 1390-000449, 1390-000450, 1390-000451, 1390-000452, 1390-000453, 1390-000454, 1390-000455, 1390-000456, 1390-000457, 1390-000458, 1390-000459, 1390-000460, 1390-000461, 1390-000462, 1390-000463, 1390-000464, 1390-000465, 1390-000467, 1390-000468, 1390-000469, 1390-000470, 1390-000483, 1390-000564, 1390-000582, 1390-000583, 1390-000584, 1390-000585, 1390-000586, 1390-000587, 1390-000588, 1390-000589, 1390-000590, 1390-000591, 1390-000592, 1390-000593, 1390-000594, 1390-000595, 1390-000597, 1390-000598, 1390-000599, 1390-000600, 1390-000. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Spacelabs Healthcare has learned through several reports that the input circuit may fail and the monitor will go on battery mode. The monitor will function normally until the batteries become depleted and then the monitor will shut down. In addition the monitor may fail to turn on due to an internal short.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Spacelabs Healthcare, Llc
- Distributor
- Worldwide Distribution.On 4/16/13 the firm advised FDA that they added serial number 1390-000055 and 1390-000913. The two additional serial numbers were added because they were provided as replacements to a French customer whose devices had been returned to support the firm's investigation. The customer had already received the Urgent Medical Device correction, customer letter and required Spacelabs to provide these devices with the understanding that they would require the field corrective action
- Lot
- Serial numbers for devices distributed in the US: 1390-000102, 1390-000103, 1390-000104, 1390-000105, 1390-000106, 1390-000107, 1390-000108, 1390-000109, 1390-000110, 1390-000111, 1390-000112, 1390-000113, 1390-000114, 1390-000115, 1390-000116, 1390-000119, 1390-000120, 1390-000150, 1390-000151, 1390-000152, 1390-000153, 1390-000154, 1390-000155, 1390-000156, 1390-000157, 1390-000158, 1390-000159, 1390-000160, 1390-000161, 1390-000162, 1390-000163, 1390-000164, 1390-000165, 1390-000166, 1390-000167, 1390-000168, 1390-000169, 1390-000170, 1390-000171, 1390-000172, 1390-000173, 1390-000174, 1390-000175, 1390-000176, 1390-000177, 1390-000178, 1390-000179, 1390-000180, 1390-000181, 1390-000182, 1390-000183, 1390-000184, 1390-000185, 1390-000186, 1390-000187, 1390-000188, 1390-000189, 1390-000190, 1390-000191, 1390-000192, 1390-000193, 1390-000195, 1390-000196, 1390-000197, 1390-000198, 1390-000199, 1390-000200, 1390-000201, 1390-000202, 1390-000203, 1390-000204, 1390-000205, 1390-000206, 1390-000207, 1390-000208, 1390-000209, 1390-000210, 1390-000211, 1390-000212, 1390-000213, 1390-000214, 1390-000215, 1390-000216, 1390-000217, 1390-000218, 1390-000219, 1390-000220, 1390-000221, 1390-000222, 1390-000223, 1390-000224, 1390-000225, 1390-000226, 1390-000227, 1390-000228, 1390-000229, 1390-000230, 1390-000231, 1390-000232, 1390-000233, 1390-000234, 1390-000235, 1390-000236, 1390-000237, 1390-000238, 1390-000239, 1390-000240, 1390-000241, 1390-000242, 1390-000243, 1390-000244, 1390-000245, 1390-000246, 1390-000247, 1390-000248, 1390-000249, 1390-000250, 1390-000251, 1390-000252, 1390-000253, 1390-000254, 1390-000255, 1390-000256, 1390-000257, 1390-000258, 1390-000259, 1390-000260, 1390-000261, 1390-000262, 1390-000263, 1390-000264, 1390-000266, 1390-000267, 1390-000268, 1390-000269, 1390-000270, 1390-000271, 1390-000272, 1390-000273, 1390-000274, 1390-000275, 1390-000276, 1390-000277, 1390-000278, 1390-000279, 1390-000280, 1390-000281, 1390-000282, 1390-000283, 1390-000284, 1390-000285, 1390-000286, 1390-000287, 1390-000288, 1390-000289, 1390-000296, 1390-000297, 1390-000298, 1390-000299, 1390-000300, 1390-000301, 1390-000302, 1390-000304, 1390-000305, 1390-000307, 1390-000308, 1390-000310, 1390-000311, 1390-000312, 1390-000314, 1390-000315, 1390-000316, 1390-000317, 1390-000318, 1390-000319, 1390-000321, 1390-000322, 1390-000323, 1390-000325, 1390-000341, 1390-000342, 1390-000343, 1390-000344, 1390-000345, 1390-000346, 1390-000347, 1390-000348, 1390-000349, 1390-000350, 1390-000351, 1390-000352, 1390-000353, 1390-000354, 1390-000355, 1390-000356, 1390-000357, 1390-000358, 1390-000359, 1390-000360, 1390-000361, 1390-000362, 1390-000363, 1390-000364, 1390-000365, 1390-000366, 1390-000367, 1390-000368, 1390-000369, 1390-000370, 1390-000371, 1390-000372, 1390-000373, 1390-000374, 1390-000375, 1390-000376, 1390-000377, 1390-000378, 1390-000379, 1390-000380, 1390-000381, 1390-000382, 1390-000383, 1390-000384, 1390-000385, 1390-000386, 1390-000387, 1390-000388, 1390-000389, 1390-000390, 1390-000391, 1390-000392, 1390-000393, 1390-000394, 1390-000420, 1390-000421, 1390-000422, 1390-000423, 1390-000424, 1390-000425, 1390-000426, 1390-000427, 1390-000428, 1390-000429, 1390-000430, 1390-000431, 1390-000432, 1390-000433, 1390-000434, 1390-000435, 1390-000436, 1390-000437, 1390-000438, 1390-000439, 1390-000440, 1390-000441, 1390-000442, 1390-000443, 1390-000444, 1390-000445, 1390-000446, 1390-000447, 1390-000448, 1390-000449, 1390-000450, 1390-000451, 1390-000452, 1390-000453, 1390-000454, 1390-000455, 1390-000456, 1390-000457, 1390-000458, 1390-000459, 1390-000460, 1390-000461, 1390-000462, 1390-000463, 1390-000464, 1390-000465, 1390-000467, 1390-000468, 1390-000469, 1390-000470, 1390-000483, 1390-000564, 1390-000582, 1390-000583, 1390-000584, 1390-000585, 1390-000586, 1390-000587, 1390-000588, 1390-000589, 1390-000590, 1390-000591, 1390-000592, 1390-000593, 1390-000594, 1390-000595, 1390-000597, 1390-000598, 1390-000599, 1390-000600, 1390-000
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASpacelabs Healthcare G2 Clinical Access Mislabeled CDs for PN: 063-1829-11 Rev A and Update Kit, PN:April 2, 2014
- FDAARIATELE TELEMETRY TRANSMITTER, Model 96281, with SpO2 Option C. Model 96281 is intended for use with eitherFebruary 5, 2014
- FDASpacelabs Healthcare qube Compact Patient Monitor, Model 91390. It is a patient monitor with arrhythmiaNovember 27, 2013
- FDASpacelabs Healthcare elance Vital Signs Monitor, Model 93300, with Option S. It is a patient monitor withNovember 20, 2013
- FDAFreedom Cordless LED Light System. Freedom System, Standard Length, Part Number 910863-1, Serial NumbersNovember 28, 2012
- FDAINTRABEAM Balloon Applicator Set; 3.0 cm, Carl Zeiss Meditec AG, Oberkochen, Germany INTRABEAM System isNovember 28, 2012
Questions about this recall
Was qube Compact Monitor (Patient Monitor) 91390. The monitor is 26.2 cm x 31.5 cm x 13.2 cm and weighs 4.1 kg (10.3 inches x 12.4 inches x 5.2 inches - 9 lbs). It has a 26.2 cm (12.1 inches) LCD display with 1024 x 768 resolution. The Spacelabs Healthcare Qube Compact Monitor (91390), functioning as either bedside or central monitors; passively displays data generated by 2 Spacelabs Healthcare parameter modules, Flexports interfaces, and other SDLC based products in the form of waveform and numeric displays, trends and alarms. Key monitored parameters available on the Model 91390 when employing the Spacelabs Command Module, consist of ECU, respiration, invasive and noninvasive blood pressure, Sp02, temperature and cardiac output. Additional parameters and interfaces to other systems are also available depending on the parameter modules employed. The Qube is intended to alert the user to alarm conditions that are reported by Spacelabs Healthcare parameter modules and/or other physiologic monitors via Flexport interfaces. These devices determine a) when an alarm condition is violated; b) the alarm priority (i.e. high, medium or low); c) alarm limits; and d) when to initiate and terminate alarm notifications. The patient monitors are also capable of displaying alarm conditions on other monitors that are on the network through the Alarm Watch feature. The Qube may also function as a generic display or computer terminal. As a generic display or terminal, the patient monitors allow network-based applications to open windows and display information on other networked monitors. The Qube is also designed to communicate with a variety of external devices such as displays, network devices, serial devices, user input devices, audio systems, and local/remote recorders. The Qube is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a hospital environment. recalled?
Yes. FDA published recall Z-0403-2013 on November 28, 2012. Reported hazard: Spacelabs Healthcare has learned through several reports that the input circuit may fail and the monitor will go on battery mode. The monitor will function normally until the batteries become depleted and then the monitor will shut down. In addition the monitor may fail to turn on due to an internal short.
What should I do if I have this product?
Voluntary: Firm initiated · Letter