The brand name of the device is Laser Loupes, a dentalloupe with laser eye protection. Laser Loupes have been assigned product code 79 FSP and are classified as a Class I Medical Devices. Laser Loupes are available with four (4) different filters of different colors. Each colored filter is associated with specific laser frequency ranges. The device listing number for this product is E160359. The intended use of this device is to magnify the operatory field and provide laser eye protection.
Published November 28, 2012 · Recall initiated: April 10, 2012
Summary
FDA records medical device Class II recall Z-0414-2013, published November 28, 2012 and initiated April 10, 2012. Recalling firm: Kerr Corporation; Status: Terminated; Reported quantity: 639 units; Lot: The Laser Loupes are custom made-to-order products. There are no general model, catalog, code, lot or serial numbers. Laser Loupes have no expiration date. The affected units were produced and distributed since September 2008. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The firm initiated the recall for Laser Loupes because the ink used to mark which laser frequency each laser Loupe protects against may wipe off. The loupes work according to specification, but the marking may not permanently adhere to the device.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Kerr Corporation
- Distributor
- Worldwide Distribution--USA (nationwide) and the countries of Canada and EU.
- Lot
- The Laser Loupes are custom made-to-order products. There are no general model, catalog, code, lot or serial numbers. Laser Loupes have no expiration date. The affected units were produced and distributed since September 2008.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDATytin Regular Set, Double Spill, 50 Capsules, Part Number 29948, Lot Number 3-1294. The intended use of thisJune 25, 2014
- FDAFreedom Cordless LED Light System. Freedom System, Standard Length, Part Number 910863-1, Serial NumbersNovember 28, 2012
- FDABioPlant, a dental bone grafting material. BioPlant has been assigned the product code LYC (Bone GraftingNovember 28, 2012
- FDACapLOX II Final Torque Driver, Indicator is a Pedicle screw spinal system. The CapLOX II Spinal System is aNovember 28, 2012
- FDARanawat/Burnstein Acetabular Shell 54mm/Limited Holes/Porous Coated is a metallic, full-hemisphere, acetabularNovember 28, 2012
- FDAThe brand name of the device is AlgiNot, a dental impression material. Product Part Numbers AlgiNot Intra KitNovember 14, 2012
Questions about this recall
Was The brand name of the device is Laser Loupes, a dentalloupe with laser eye protection. Laser Loupes have been assigned product code 79 FSP and are classified as a Class I Medical Devices. Laser Loupes are available with four (4) different filters of different colors. Each colored filter is associated with specific laser frequency ranges. The device listing number for this product is E160359. The intended use of this device is to magnify the operatory field and provide laser eye protection. recalled?
Yes. FDA published recall Z-0414-2013 on November 28, 2012. Reported hazard: The firm initiated the recall for Laser Loupes because the ink used to mark which laser frequency each laser Loupe protects against may wipe off. The loupes work according to specification, but the marking may not permanently adhere to the device.
What should I do if I have this product?
Voluntary: Firm initiated · Letter