KING LAD SILI CONE SINGLE USE, King Systems, 10 devices/case, ETO Sterile Size 3 / Child Silicone King LAD® Patient Size: 30-50kg Product Usage: Laryngeal airway
Published November 28, 2012 · Recall initiated: November 2, 2012
Summary
FDA records medical device Class II recall Z-0419-2013, published November 28, 2012 and initiated November 2, 2012. Recalling firm: King Systems Corp.; Status: Terminated; Reported quantity: 263 cases of product. (2,630 devices); Lot: REF: LAD-303, Lot #K120618. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The firm received two complaints from a user facility in New Jersey regarding color-coded EtO sterilization indicators on the primary device packaging that did not undergo the expected color change. The complainant reported devices that had pink color indicators when the product packaging specified that the color should change from pink to yellow (or brown) if the product had subjected to an EtO
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- King Systems Corp.
- Distributor
- US Nationwide Distribution - WI, TX, NY, MO, IN, KY, FL, NJ, PA, TN, and NC
- Lot
- REF: LAD-303, Lot #K120618
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAKing Systems KLTSD414 Quantity, 5 PER CASE KING LTSD KIT SIZE 4 King LT(S)-DTM supraglottic airwayJanuary 22, 2014
- FDAUniversal Flex2 Breathing Circuit Class I 510(k) Exempt An anesthesia breathing circuit is a device that isOctober 9, 2013
- FDAKing LTS-D Oropharyngeal Airways Product Usage: The King LTS-D is intended for use in adult patients (inOctober 9, 2013
- FDAKing LT-D Oropharyngeal Airways The King LT-D is intended for use in adult patients (in excess of 25 kg) forSeptember 25, 2013
- FDAKING LAD SILI CONE SINGLE USE, King Systems devices/case, ETO Sterile. Size 2 / Infant Silicone King LAD®November 28, 2012
- FDAKING LAD SILI CONE SINGLE USE, King Systems, 10 devices/case, ETO Sterile Size 4 / Adult Silicone King LAD®November 28, 2012
Questions about this recall
Was KING LAD SILI CONE SINGLE USE, King Systems, 10 devices/case, ETO Sterile Size 3 / Child Silicone King LAD® Patient Size: 30-50kg Product Usage: Laryngeal airway recalled?
Yes. FDA published recall Z-0419-2013 on November 28, 2012. Reported hazard: The firm received two complaints from a user facility in New Jersey regarding color-coded EtO sterilization indicators on the primary device packaging that did not undergo the expected color change. The complainant reported devices that had pink color indicators when the product packaging specified that the color should change from pink to yellow (or brown) if the product had subjected to an EtO
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail