Siemens Dimension HB1C Flex reagent cartridge; Dimension Hemoglobin A1c (HB1C) Flex reagent cartridge on the Dimension clinical chemistry system is an in vitro diagnostic assay for the quantitative determination of hemoglobin A1c (HbA1c) in human anticoagulated whole blood. Measurements of hemoglobin A1c are effective in monitoring long-term glucose control in individuals with diabetes mellitus.
Published December 3, 2014 · Recall initiated: September 30, 2014
Summary
FDA records medical device Class II recall Z-0420-2015, published December 3, 2014 and initiated September 30, 2014. Recalling firm: Siemens Healthcare Diagnostics, Inc.; Status: Terminated; Reported quantity: 53, 046; Lot: Siemens Dimension Vista HbA1c Flex reagent cartridge: Catalog Number K3105A, Siemens Material Number (SMN) 10470481, Lot Numbers 13294BA 13308BA 13329AA 13350AA 14006AA 14041BA 14069BB 14083BA Siemens Dimension HB1C Flex reagent cartridge: Catalog Number DF105A, Siemens Material Number (SMN) 10483822, Lot Numbers GA4266 BA4273 BA4280 BA4287 GA4301 GA4315 GC4322 GA4343 GA4350 GA4357 GA5013 GA5020. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Certain lots of the reagent cartridges show confirmed positive bias averaging 0.4% Hemoglobin A1c units and occasionally up to 1.0% HbA1c units for patient samples when compared to the National Glycohemoglobin Standardization Program (NGSP). QC samples may exhibit a similar bias.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Distributor
- Worldwide Distribution: US (nationwide) and internationally to: Canada.
- Lot
- Siemens Dimension Vista HbA1c Flex reagent cartridge: Catalog Number K3105A, Siemens Material Number (SMN) 10470481, Lot Numbers 13294BA 13308BA 13329AA 13350AA 14006AA 14041BA 14069BB 14083BA Siemens Dimension HB1C Flex reagent cartridge: Catalog Number DF105A, Siemens Material Number (SMN) 10483822, Lot Numbers GA4266 BA4273 BA4280 BA4287 GA4301 GA4315 GC4322 GA4343 GA4350 GA4357 GA5013 GA5020
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAtellica CH Revised C-Reactive Protein (RCRP); Siemens Material Number (SMN): 11537223;October 7, 2026
- FDABeckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mLAugust 26, 2026
- FDASyva EMIT 2000 Quinidine Assay REF: 4Q019UL Contents: 1x28 mL, 1x14mLAugust 26, 2026
- FDASyva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219ULAugust 12, 2026
- FDAGyrus ACMI Disposable Falope-Ring Band Applicator Kits. REF 006889-901 and REF 005280-901 (with 8 mmDecember 3, 2014
- FDAGemStar infusion pump, list numbers 13000 and 13100, GemStar Infusion Pump (pain management) list numberDecember 3, 2014
Questions about this recall
Was Siemens Dimension HB1C Flex reagent cartridge; Dimension Hemoglobin A1c (HB1C) Flex reagent cartridge on the Dimension clinical chemistry system is an in vitro diagnostic assay for the quantitative determination of hemoglobin A1c (HbA1c) in human anticoagulated whole blood. Measurements of hemoglobin A1c are effective in monitoring long-term glucose control in individuals with diabetes mellitus. recalled?
Yes. FDA published recall Z-0420-2015 on December 3, 2014. Reported hazard: Certain lots of the reagent cartridges show confirmed positive bias averaging 0.4% Hemoglobin A1c units and occasionally up to 1.0% HbA1c units for patient samples when compared to the National Glycohemoglobin Standardization Program (NGSP). QC samples may exhibit a similar bias.
What should I do if I have this product?
Voluntary: Firm initiated · Letter