EZ Glide Aortic Cannula, 24 Fr. (8.0mm) x 37.6 cm (14.8") 3/8" (9.5 mm) vented connector , RX Only, Sterile EC, Made in Mexico, UDI: 00690103172119 for the following REF #s (UDI): EZC21A (00690103174182), EZC21TA (00690103174199), EZC24A (00690103714401), EZC24TA (00690103714418), EZF21A (00690103175080), EZF21TA (00690103175097), EZF24A (00690103175066), EZF24TA (00690103175073), EZS21A (00690103172096), EZS21TA (00690103172102), EZS24A (00690103172119), and EZS24TA (00690103172126) Product Usage: Aortic perfusion cannulae are intended for perfusion of the ascending aorta during short-term (< 6 hours) cardiopulmonary bypass procedures. Aortic cannulae in sizes 6 Fr. (2 mm) to 18 Fr. (6 mm) can be used in pediatric patient populations.
Published November 27, 2019 · Recall initiated: October 22, 2019
Summary
FDA records medical device Class II recall Z-0420-2020, published November 27, 2019 and initiated October 22, 2019. Recalling firm: Edwards Lifesciences, LLC; Status: Terminated; Reported quantity: 62,431 units; Lot: Lot #s: 203876 207654 217024 226788,205724, 207655, 217025, 226938,205775, 208823, 217048, 227494,205868, 208893, 218185, 227886,205920, 208894, 218457, 228009,205921, 209901, 218539, 228259,205922, 209904, 219090, 228549,206369, 210110, 219212, 228558,206372, 210111, 219561, 229051,206373, 212032 ,220146 ,229054,206374, 212033 ,220151 ,229827,206768, 212034 ,220157 ,230690,206769, 212035 ,220802 ,230894,206770, 212258 ,220803 ,231105,206771, 212280 ,222841 ,231109,206772, 212959 ,222913 ,231111,206773, 213506 ,222914 ,231291,206774, 213914 ,222915 ,231485,206775, 214214 ,223174 ,232109,206776 ,214216 ,224233 ,232437,206777 ,214716 ,224235 ,234138,206778 ,214717 ,224662,207319 ,214961 ,224707,207320 ,215239 ,225306,207321 ,215495 ,225884,207322 ,216123 ,225885,207323 ,216196 ,226188,207324 ,216627 ,226596,207636 ,216628 ,226597,207638 ,216632, 226691. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Their is a possibility that the cannula may separate from its connector, potentially causing a breach of circuit during cardiopulmonary bypass which could result in significant blood loss.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Edwards Lifesciences, LLC
- Distributor
- Worldwide distribution - US Nationwide distribution in the states of Alaska, Alabama, Arkansas, Arizona, California, Colorado, Connecticut, Washington DC, Delaware, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minnesota, Missouri, Mississippi, Montana, North Carolina, North Dakota, Nebraska, New Hampshire, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, Tennessee, Texas, Utah, Virginia, Vermont, Washington, Washington DC, Wisconsin, West Virginia, Wyoming and countries of Austria, Australia, Belgium, Bahrain, Canada, Switzerland, Chile, China, Cyprus, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, Hungary, Canary Islands, Ireland, India, Iceland, Italy, Japan, Mauritius, Malaysia, Netherlands, Norway, Poland, Romania, Sweden, Singapore, Slovakia, Thailand, Turkey, United Arab Emirates, South Africa.
- Lot
- Lot #s: 203876 207654 217024 226788,205724, 207655, 217025, 226938,205775, 208823, 217048, 227494,205868, 208893, 218185, 227886,205920, 208894, 218457, 228009,205921, 209901, 218539, 228259,205922, 209904, 219090, 228549,206369, 210110, 219212, 228558,206372, 210111, 219561, 229051,206373, 212032 ,220146 ,229054,206374, 212033 ,220151 ,229827,206768, 212034 ,220157 ,230690,206769, 212035 ,220802 ,230894,206770, 212258 ,220803 ,231105,206771, 212280 ,222841 ,231109,206772, 212959 ,222913 ,231111,206773, 213506 ,222914 ,231291,206774, 213914 ,222915 ,231485,206775, 214214 ,223174 ,232109,206776 ,214216 ,224233 ,232437,206777 ,214716 ,224235 ,234138,206778 ,214717 ,224662,207319 ,214961 ,224707,207320 ,215239 ,225306,207321 ,215495 ,225884,207322 ,216123 ,225885,207323 ,216196 ,226188,207324 ,216627 ,226596,207636 ,216628 ,226597,207638 ,216632, 226691
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAGeneric Name: Edwards Commander M Delivery System Generic Name: Edwards Commander M Delivery SystemNovember 11, 2026
- FDAEdwards EVOQUE tricuspid delivery system REF:9850TDS Sterile EOMay 20, 2026
- FDABrand Name: Optisite Arterial Cannula Product Name: Optisite Arterial Cannula Model/Catalog Number: OPTI16July 16, 2025
- FDABrand Name: FEM-FLEX II Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII016AS SoftwareJuly 16, 2025
- FDACoolBlue Hyper-Hypothermia System- body hypothermia system Model: Innercool - Product Usage: The system isNovember 27, 2019
- FDATorcon NB Advantage Catheters - Product Usage: Intended for use in the peripheral and coronary vascularNovember 27, 2019
Questions about this recall
Was EZ Glide Aortic Cannula, 24 Fr. (8.0mm) x 37.6 cm (14.8") 3/8" (9.5 mm) vented connector , RX Only, Sterile EC, Made in Mexico, UDI: 00690103172119 for the following REF #s (UDI): EZC21A (00690103174182), EZC21TA (00690103174199), EZC24A (00690103714401), EZC24TA (00690103714418), EZF21A (00690103175080), EZF21TA (00690103175097), EZF24A (00690103175066), EZF24TA (00690103175073), EZS21A (00690103172096), EZS21TA (00690103172102), EZS24A (00690103172119), and EZS24TA (00690103172126) Product Usage: Aortic perfusion cannulae are intended for perfusion of the ascending aorta during short-term (< 6 hours) cardiopulmonary bypass procedures. Aortic cannulae in sizes 6 Fr. (2 mm) to 18 Fr. (6 mm) can be used in pediatric patient populations. recalled?
Yes. FDA published recall Z-0420-2020 on November 27, 2019. Reported hazard: Their is a possibility that the cannula may separate from its connector, potentially causing a breach of circuit during cardiopulmonary bypass which could result in significant blood loss.
What should I do if I have this product?
Voluntary: Firm initiated · Letter