Back to search
FDAFDA · Food and drugs·Recall number Z-0424-2021

Covidien MONOJECT Standard Hypodermic Needle, Polypropylene Hub 22 G x 3/4". Intended use to facilitate activities such as: withdrawal of medication from a container, aspiration of fluid from a patient, injection of medication into a patient, and transfer fluids Item Code: 8881250248

Published November 11, 2020 · Recall initiated: September 25, 2020

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Potential for the needle to detach from the cartridge after the bottom of the cartridge is twisted and removed. If the needle is not securely seated, the needle can fall out of the sheath and cause the unused needle to project out of the cartridge, and result in a needle stick/puncture from the clean/unused needle

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Cardinal Health 200, LLC
Distributor
Distribution US Nationwide, Australia and Canada
Lot
Lot Number: 011144 Expanded Recall 11/4/20: Lot Number: 011145
Risk classification
Class II
Area (FDA)
device
Country
United States