FDAFDA · Food and drugs·Recall number Z-0424-2021
Covidien MONOJECT Standard Hypodermic Needle, Polypropylene Hub 22 G x 3/4". Intended use to facilitate activities such as: withdrawal of medication from a container, aspiration of fluid from a patient, injection of medication into a patient, and transfer fluids Item Code: 8881250248
Published November 11, 2020 · Recall initiated: September 25, 2020
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential for the needle to detach from the cartridge after the bottom of the cartridge is twisted and removed. If the needle is not securely seated, the needle can fall out of the sheath and cause the unused needle to project out of the cartridge, and result in a needle stick/puncture from the clean/unused needle
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Cardinal Health 200, LLC
- Distributor
- Distribution US Nationwide, Australia and Canada
- Lot
- Lot Number: 011144 Expanded Recall 11/4/20: Lot Number: 011145
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States