FDAFDA · Food and drugs·Recall number Z-0440-2014
Navigator HD l3F/15F x 28 cm Ureteral Access Sheath Set Product No M0062502270, RDF Catalog No 250-227, Sterile. Indicated for use in endoscopic procedures to facilitate the passage of endoscopes, urological instruments and for the injection of fluids into the urinary tract.
Published December 11, 2013 · Recall initiated: November 5, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The 13F/15F Navigator HD Ureteral Access SheathSet are being recalled after receipt of 5 reports where the tip of the dilator has separated fro the body of the dilator.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Boston Scientific Corporation
- Distributor
- Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Ireland, Denmark, Italy, France, Canada, Great Britain, Switzerland, South Africa, Germany, and the Netherlands.
- Lot
- UPN M0062502270, Catalog 250-227, Lot 15708431, 15708432, 15708433, 15812875, 15836355, 15855545, 15876213 , 15896950, 15935564, 16028474, 16044445, and 16046280
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Sell products like this?
Recalert watches your whole catalog for recalls like this one, automatically.