iDesign, Model - G300: System AWS (International), System AWS (China), Advanced WaveScan Studio (United States); iDesign Refractive Studio Aberrometer (International) and (United States), Model: G301
Published January 5, 2022 · Recall initiated: October 21, 2021
Summary
FDA records medical device Class II recall Z-0443-2022, published January 5, 2022 and initiated October 21, 2021. Recalling firm: AMO Manufacturing USA, LLC; Status: Ongoing; Reported quantity: 711; Lot: G300 with Windows XP. G301 with Windows 7 Embedded System. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Aberrometer and precision laser systems are affected by a remote code execution vulnerability existing when the Microsoft Windows Print Spooler service is enabled, and improperly performs privileged file operations. This could be exploited to run arbitrary code with SYSTEM privileges, allowing for program installation; view/change/deleting data; or creation of new accounts with full user rights.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- AMO Manufacturing USA, LLC
- Distributor
- US: CA, TX, FL, LA, MD, IA, MA, VA, SD, MN, MO, NC, TN, SC, WA, NV, OH, KY, IN, CO, NJ, OK, NH, UT, AR, NY, AK, ND, NE, IL, AL, HI, CT, DE, KS, GA, OR, MS, PA, MT, ME, MI, AZ, ID, WI, VT, NM, CA. OUS: Canada, Mexico, Russian Federation, Turkey, Brazil, Egypt, Iraq, Saudi Arabia, Jordan, Norway, Colombia, Germany, Malaysia, India, Oman, Azerbaijan, Canada, Slovakia, Lebanon, Spain, Ireland, Italy, South Korea, Bangladesh, Chile, Taiwan, Tunisia, Kazakhstan, Libya, Argentina, Singapore, Austria, Australia, Vietnam, Greece, Poland, Indonesia, Yemen, Great Britain, South Africa, Croatia, France, Czech Republic, Algeria, Thailand, Bulgaria, United Arab Emirates, Belgium, Japan, China, Nepal, Israel, Switzerland, Dominican Republic, Hong Kong, Portugal, Ecuador, Kuwait, Netherlands, Slovenia, Philippines
- Lot
- G300 with Windows XP. G301 with Windows 7 Embedded System.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAGatifloxacin Ophthalmic Solution, 0.5%, 2.5 mL bottle, Rx only, Manufactured by: Lupin Limited, PithampurJanuary 12, 2022
- FDAPAIN AID ESF- (Acetaminopehn USP 250mg, Asprin USP 250mg Caffeine 65mg) coated, bulk OTC tablets packaged inJanuary 12, 2022
- FDAEQUALINE aller-ease fexofenadine hydrochloride tablets, 180 mg, 24 HR, packaged as a) 15 count bottle, NDCJanuary 12, 2022
- FDAARTISTE with syngo RT Therapist-linear accelerator systems to deliver X-Ray photon and electron radiation forJanuary 5, 2022
- FDAVIDAS RUB IgG (RBG), REF 30226January 5, 2022
- FDAGiant Eagle, Diced Green Peppers, 10 oz. poly bags and 22# cases, also poly bags and cartons, 12/10 oz. bagsJanuary 5, 2022
Questions about this recall
Was iDesign, Model - G300: System AWS (International), System AWS (China), Advanced WaveScan Studio (United States); iDesign Refractive Studio Aberrometer (International) and (United States), Model: G301 recalled?
Yes. FDA published recall Z-0443-2022 on January 5, 2022. Reported hazard: Aberrometer and precision laser systems are affected by a remote code execution vulnerability existing when the Microsoft Windows Print Spooler service is enabled, and improperly performs privileged file operations. This could be exploited to run arbitrary code with SYSTEM privileges, allowing for program installation; view/change/deleting data; or creation of new accounts with full user rights.
What should I do if I have this product?
Voluntary: Firm initiated · Letter