Kimberly-Clark Patient Warming System - Model 1000 Control Unit The product is a thermal regulating system, indicated for monitoring and controlling patient temperature.
Published December 5, 2012 · Recall initiated: September 16, 2010
Summary
FDA records medical device Class II recall Z-0451-2013, published December 5, 2012 and initiated September 16, 2010. Recalling firm: Kimberly-Clark Corporation; Status: Terminated; Reported quantity: 400 units; Lot: Serial Number - SA09043017, SA09043019, SA0904320, SA09043018, SA09043015, SA09043012, SA09043016, SA09043014, SA09043013, SA09043009, SA09043011, SA09043010, SA09043008, SA09043007, SA09043006, SA09043003, SA09043005, SA09043004, SA09043002, SA09043001, SA09020013, SA09020012, SA09020014, SA09020011, SA09020010, SA09020009, SA0902008, SA09020006, SA09020007, SA08343100, SA09020004, SA09020005, SA09020002, SA09020003, SA09020001, SA08343099, SA08343098, SA08343097, SA08343096, SA08343095, SA08343094, SA08343093, SA08343092, SA08343091, SA08343090, SA08343089, SA08343087, SA08343088, SA08343086, SA08343085, SA08343083, SA08343084, SA08343053, SA08343073, SA08288014, SA08288002, SA08343054, SA08343082, SA08343081, SA08343080, SA08343079, SA08343078, SA08343077, SA08165007, SA08343076, SA08343075, SA08343074, SA08343072, SA08343055, SA08288007, SA08343071, SA08343070, SA08343069, SA08343068, SA08343067, SA08343066, SA08343065, SA08343064, SA08343063, SA08343059, SA08343058, SA08343062, SA08343061, SA08343060, SA08343056, SA08343057, SA08288012, SA08288017, SA08288016, SA08288015, SA08288013, SA08288011, SA08288009, SA08288006, SA08288003, SA08288010, SA08288008, SA08288004, SA08288005, SA08288001, SA08165013, SA08165014, SA08165015, SA08165018, SA08165016, SA08165017, SA08165006, SA08165005, SA08165011, SA08165012, SA08165010, SA08165009, SA08165004, SA08165008, SA08165002, SA08165003, SA08144035, SA08144030, SA08165001, SA08144034, SA08144033, SA08144032, SA08144031, SA08144029, SA08144028, SA08144025, SA08144027, SA08144021, SA08144026, SA08144024, SA08144023, SA08144022, SA08136010, SA08136006, SA08136003, SA08136009, SA08136005, SA08136004, SA08136007, SA08136001, SA08136002, SA08136008, SA08094129, SA08094137, SA08094141, SA08094136, SA08094150, SA08094153, SA08094152, SA08094151, SA08094149, SA08094145, SA08094148, SA08094147, SA08094143, SA08094146, SA08094142, SA08094144, SA08094140, SA08094139, SA08094138, SA08094134, SA08094130, SA08094132, SA08094135, SA08094131, SA08094133, SA08094127, SA08094128, SA08094125, SA08094126, SA08094124, SA08017044, SA08017049, SA08017045, SA08017040, SA08017034, SA08017035, SA08017041, SA08017047, SA08017042, SA08017043, SA08017048, SA08017039, SA08017046, SA08017038, SA08017037, SA08017030, SA08017036, SA08017028, SA08017029, SA08017027, SA08017033, SA08017031, SA08017032, SA07344022, SA07344029, SA07344028, SA07344026, SA07344027, SA07344030, SA07344021, SA07344023, SA07344024, SA07344025, SA07320002, SA07320001, SA07320014, SA07320015, SA07320013, SA07320011, SA07320012, SA07320006, SA07320005, SA07320008, SA07320010, SA07320009, SA07320007, SA07320004, SA07320003, SA07261006, SA07257008, SA07261004, SA07261007, SA07261002, SA07257004, SA07257007, SA07257003, SA07257002, SA07261005, SA07261008, SA07261003, SA07261001, SA07257005, SA07257006, SA07257001, SA07250005, SA07250004, SA07250002, SA07250003, SA07250001, SA07228007, SA07228008, SA07228001, SA07228002, SA07228003, SA07228004, SA07228005, SA07228006, SA07180006, SA07180005, SA07180004, SA07180003, SA07180002, SA07180001, SC07171006, SC07171005, SC07171004, SC07171003, SC07171002, SC07171001. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Error 85 alarm on the Model 1000 Control Unit was occasionally triggered inappropriately.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Kimberly-Clark Corporation
- Distributor
- Nationwide Distribution including AZ, AK, CA, CO, DE, DC, FL, GA. ID, IL, IN, IA, KS, KY, LA, MD, ME MA, MS, MO, MT, NE, NJ, NM, NV, NY, NC, OH, OR, PA, RI, SC, TN, TX, VA, WV, and WI
- Lot
- Serial Number - SA09043017, SA09043019, SA0904320, SA09043018, SA09043015, SA09043012, SA09043016, SA09043014, SA09043013, SA09043009, SA09043011, SA09043010, SA09043008, SA09043007, SA09043006, SA09043003, SA09043005, SA09043004, SA09043002, SA09043001, SA09020013, SA09020012, SA09020014, SA09020011, SA09020010, SA09020009, SA0902008, SA09020006, SA09020007, SA08343100, SA09020004, SA09020005, SA09020002, SA09020003, SA09020001, SA08343099, SA08343098, SA08343097, SA08343096, SA08343095, SA08343094, SA08343093, SA08343092, SA08343091, SA08343090, SA08343089, SA08343087, SA08343088, SA08343086, SA08343085, SA08343083, SA08343084, SA08343053, SA08343073, SA08288014, SA08288002, SA08343054, SA08343082, SA08343081, SA08343080, SA08343079, SA08343078, SA08343077, SA08165007, SA08343076, SA08343075, SA08343074, SA08343072, SA08343055, SA08288007, SA08343071, SA08343070, SA08343069, SA08343068, SA08343067, SA08343066, SA08343065, SA08343064, SA08343063, SA08343059, SA08343058, SA08343062, SA08343061, SA08343060, SA08343056, SA08343057, SA08288012, SA08288017, SA08288016, SA08288015, SA08288013, SA08288011, SA08288009, SA08288006, SA08288003, SA08288010, SA08288008, SA08288004, SA08288005, SA08288001, SA08165013, SA08165014, SA08165015, SA08165018, SA08165016, SA08165017, SA08165006, SA08165005, SA08165011, SA08165012, SA08165010, SA08165009, SA08165004, SA08165008, SA08165002, SA08165003, SA08144035, SA08144030, SA08165001, SA08144034, SA08144033, SA08144032, SA08144031, SA08144029, SA08144028, SA08144025, SA08144027, SA08144021, SA08144026, SA08144024, SA08144023, SA08144022, SA08136010, SA08136006, SA08136003, SA08136009, SA08136005, SA08136004, SA08136007, SA08136001, SA08136002, SA08136008, SA08094129, SA08094137, SA08094141, SA08094136, SA08094150, SA08094153, SA08094152, SA08094151, SA08094149, SA08094145, SA08094148, SA08094147, SA08094143, SA08094146, SA08094142, SA08094144, SA08094140, SA08094139, SA08094138, SA08094134, SA08094130, SA08094132, SA08094135, SA08094131, SA08094133, SA08094127, SA08094128, SA08094125, SA08094126, SA08094124, SA08017044, SA08017049, SA08017045, SA08017040, SA08017034, SA08017035, SA08017041, SA08017047, SA08017042, SA08017043, SA08017048, SA08017039, SA08017046, SA08017038, SA08017037, SA08017030, SA08017036, SA08017028, SA08017029, SA08017027, SA08017033, SA08017031, SA08017032, SA07344022, SA07344029, SA07344028, SA07344026, SA07344027, SA07344030, SA07344021, SA07344023, SA07344024, SA07344025, SA07320002, SA07320001, SA07320014, SA07320015, SA07320013, SA07320011, SA07320012, SA07320006, SA07320005, SA07320008, SA07320010, SA07320009, SA07320007, SA07320004, SA07320003, SA07261006, SA07257008, SA07261004, SA07261007, SA07261002, SA07257004, SA07257007, SA07257003, SA07257002, SA07261005, SA07261008, SA07261003, SA07261001, SA07257005, SA07257006, SA07257001, SA07250005, SA07250004, SA07250002, SA07250003, SA07250001, SA07228007, SA07228008, SA07228001, SA07228002, SA07228003, SA07228004, SA07228005, SA07228006, SA07180006, SA07180005, SA07180004, SA07180003, SA07180002, SA07180001, SC07171006, SC07171005, SC07171004, SC07171003, SC07171002, SC07171001
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAScott® Moisturizing Foam Hand Sanitizer, 1.2 Liters (40.5 fl oz), Distributed in the U.S. by Kimberly-ClarkApril 14, 2021
- FDAU by Kotex® Sleek®, Regular Tampons, 3 CountJanuary 23, 2019
- FDAU by Kotex® Sleek®, Regular Tampons, 18 CountJanuary 23, 2019
- FDAKimberly-Clark RadiOpaque Radiofrequency Cannula, Gauge 22, Length 100 mm, Active Tip 10 mm, Distributed inFebruary 5, 2014
- FDANovarel® (Chorionic Gonadotropin for injection, USP), 10,000 USP FOR INTRAMUSCULAR USE ONLY Rx Only NDCDecember 12, 2012
- FDAGE Healthcare Optima XR 200 amx with Digital Upgrade. The GE Automatic Mobile X-Ray (AMX) Series: BrivoDecember 5, 2012
Questions about this recall
Was Kimberly-Clark Patient Warming System - Model 1000 Control Unit The product is a thermal regulating system, indicated for monitoring and controlling patient temperature. recalled?
Yes. FDA published recall Z-0451-2013 on December 5, 2012. Reported hazard: Error 85 alarm on the Model 1000 Control Unit was occasionally triggered inappropriately.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit