FDAFDA · Food and drugs·Recall number Z-0459-2015
PEDIATRIC Radiotransparent Electrode, REF/Catalog Number 2603H, Rx ONLY --- COMPARABLE TO PHILIPS MEDICAL HEARTSTART ELECTRODES PLUG STYLE CONNECTOR
Published December 10, 2014 · Recall initiated: November 6, 2014
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Philips Healthcare made changes to the design of the connection between multifunction electrodes and their HeartStart FR3 and HeartStart FRx AEDs. The failure to connect may result in a delay of therapy, which could result in patient death or serious injury.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- ConMed Corporation
- Distributor
- Worldwide Distribution - USA (nationwide) Puerto Rico, and Internationally to Bosnia, Herzegovina, Brazil, Canada, Great Britain, Israel, Italy, Republic of Korea, Norway, Portugal, Saudi Arabia, and Spain.
- Lot
- All Lot Codes
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Sell products like this?
Recalert watches your whole catalog for recalls like this one, automatically.