Back to search
FDAFDA · Food and drugs·Recall number Z-0459-2015

PEDIATRIC Radiotransparent Electrode, REF/Catalog Number 2603H, Rx ONLY --- COMPARABLE TO PHILIPS MEDICAL HEARTSTART ELECTRODES PLUG STYLE CONNECTOR

Published December 10, 2014 · Recall initiated: November 6, 2014

Summary

FDA records medical device Class I recall Z-0459-2015, published December 10, 2014 and initiated November 6, 2014. Recalling firm: ConMed Corporation; Status: Terminated; Reported quantity: 174,610 (168,090 US and 6,520 OUS) in total.; Lot: All Lot Codes. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Philips Healthcare made changes to the design of the connection between multifunction electrodes and their HeartStart FR3 and HeartStart FRx AEDs. The failure to connect may result in a delay of therapy, which could result in patient death or serious injury.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
ConMed Corporation
Distributor
Worldwide Distribution - USA (nationwide) Puerto Rico, and Internationally to Bosnia, Herzegovina, Brazil, Canada, Great Britain, Israel, Italy, Republic of Korea, Norway, Portugal, Saudi Arabia, and Spain.
Lot
All Lot Codes
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was PEDIATRIC Radiotransparent Electrode, REF/Catalog Number 2603H, Rx ONLY --- COMPARABLE TO PHILIPS MEDICAL HEARTSTART ELECTRODES PLUG STYLE CONNECTOR recalled?

Yes. FDA published recall Z-0459-2015 on December 10, 2014. Reported hazard: Philips Healthcare made changes to the design of the connection between multifunction electrodes and their HeartStart FR3 and HeartStart FRx AEDs. The failure to connect may result in a delay of therapy, which could result in patient death or serious injury.

What should I do if I have this product?

Voluntary: Firm initiated · Letter