ARTIS Q.zen biplane with Material 10848355 - a Angiographic x-ray system Product Usage: Artis zee is a dedicated angiography system developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis zee includes cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. Artis zee can also support the acquisition of position triggered imaging for spatial data synthesis.

Published November 21, 2018 · Recall initiated: October 19, 2018

Summary

FDA records medical device Class II recall Z-0459-2019, published November 21, 2018 and initiated October 19, 2018. Recalling firm: Siemens Medical Solutions USA, Inc; Status: Terminated; Reported quantity: 15; Lot: Serial Numbers: Serial 123202 123211 123208 123209 123206 123207 123219 123205 123201 123055 123056 123216 123200 123212. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Vibrations of the C-arm of the plane B may result in reduced image quality caused by the screw connection to the C-arm in the ceiling support

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Siemens Medical Solutions USA, Inc
Distributor
US Nationwide Distribution
Lot
Serial Numbers: Serial 123202 123211 123208 123209 123206 123207 123219 123205 123201 123055 123056 123216 123200 123212
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was ARTIS Q.zen biplane with Material 10848355 - a Angiographic x-ray system Product Usage: Artis zee is a dedicated angiography system developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis zee includes cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. Artis zee can also support the acquisition of position triggered imaging for spatial data synthesis. recalled?

Yes. FDA published recall Z-0459-2019 on November 21, 2018. Reported hazard: Vibrations of the C-arm of the plane B may result in reduced image quality caused by the screw connection to the C-arm in the ceiling support

What should I do if I have this product?

Voluntary: Firm initiated · Letter