Back to search
FDAFDA · Food and drugs·Recall number Z-0460-2025

MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment Item Number: 710001B0

Published November 20, 2024 · Recall initiated: October 3, 2024

Summary

FDA records medical device Class II recall Z-0460-2025, published November 20, 2024 and initiated October 3, 2024. Recalling firm: Getinge Usa Sales Inc; Status: Ongoing; Reported quantity: 112 units; Lot: UDI-DI: 04046768136298 S/N: 680 through 707. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Error code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control and may result in procedural delay

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Getinge Usa Sales Inc
Distributor
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bhutan, Brazil, Canada, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Ecuador, Egypt, France, Germany, Greece, India, Israel, Japan, Jordan, Korea, Republic Of Kuwait, "Macedonia, The Former Yugoslav Republic Of", Malaysia, Mauritius, Mexico, Montenegro, Netherlands, New Zealand, Panama, Peru, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Viet Nam.
Lot
UDI-DI: 04046768136298 S/N: 680 through 707
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment Item Number: 710001B0 recalled?

Yes. FDA published recall Z-0460-2025 on November 20, 2024. Reported hazard: Error code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control and may result in procedural delay

What should I do if I have this product?

Voluntary: Firm initiated · Letter