The J700 Floor Mounted Tube Stand (FMTS) is intended to support and position the diagnostic x-ray tube housing assembly for a medical radiographic procedure. The product consists of a vertical column (with counterbalanced vertical carriage, counterweight and pulley system), horizontal floor car and a floor mounted track to ride on.

Published December 3, 2014 · Recall initiated: October 28, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The welds on J700 tube stands may be insufficient resulting in tube stand breakage, damage to the X-Ray machine, and/or patient or healthcare provider injury.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Summit Industries Inc.
Distributor
US Distribution including the states of: AZ, CO, IL, KY, LA, NE, NY, OH, PA, TN, TX, VA, WA and WV.
Lot
Model Numbers: J700-01, J700-02, J700-03, J700-04
Risk classification
Class II
Area (FDA)
device
Country
United States