Enterococcus Screen Agar QUAD Plate
Published December 30, 2015 · Recall initiated: June 29, 2015
Summary
FDA records medical device Class II recall Z-0474-2016, published December 30, 2015 and initiated June 29, 2015. Recalling firm: Becton Dickinson & Co.; Status: Terminated; Reported quantity: 770; Lot: catalog number 222201. Lot numbers 5064938, 5085954, 5105509, and 5133580. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The affected expired lots of product may have exhibited breakthrough growth of vancomycin susceptible Enterococcus facaelis ATCC 29212 near the end of shelf life. The in-date product may also exhibit breakthrough growth of vancomycin susceptible Enterococcus facaelis ATCC 29212 near the end of shelf life.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Becton Dickinson & Co.
- Distributor
- Worldwide Distribution - US (nationwide) Distribution to the states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, WI and WV., and to the countries of : Canada, Colombia, India, Mexico and Europe.
- Lot
- catalog number 222201. Lot numbers 5064938, 5085954, 5105509, and 5133580
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABD Phoenix Panel NMIC-312. Catalog Number: 449082.October 21, 2026
- FDABD Phoenix Panel NMIC-306. Catalog Number: 449292.October 21, 2026
- FDABD GasPak EZ CO2 Pouch System. Catalog Number: 260684. for microbiologic useMay 6, 2026
- FDABD GasPak EZ Campy Pouch System. Catalog Number: 260685. for microbiologic useMay 6, 2026
- FDAHook Blade, Angled. Part number 7058A. Sterile disposable blade for use with Endoscopic Tarsal TunnelDecember 30, 2015
- FDA1 Step 12 Panel Cup, Item No. NBCA-12M-WDecember 30, 2015
Questions about this recall
Was Enterococcus Screen Agar QUAD Plate recalled?
Yes. FDA published recall Z-0474-2016 on December 30, 2015. Reported hazard: The affected expired lots of product may have exhibited breakthrough growth of vancomycin susceptible Enterococcus facaelis ATCC 29212 near the end of shelf life. The in-date product may also exhibit breakthrough growth of vancomycin susceptible Enterococcus facaelis ATCC 29212 near the end of shelf life.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail