Insulia Diabetes Management Companion
Published November 28, 2018 · Recall initiated: October 8, 2018
Summary
FDA records medical device Class II recall Z-0478-2019, published November 28, 2018 and initiated October 8, 2018. Recalling firm: Voluntis; Status: Terminated; Reported quantity: 14; Lot: Software version numbers 1.4.3 / 1.6.3 / 1.6.31 / 1.6.40. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The firm identified an issue with the basal calculator identified on the Android version of the Insulia application. The bug could lead from low-impact to high-impact hypoglycemia depending on the circumstances. A new update of Insulia is now available (version 1.6.42), which resolves this issue.
What to do
Voluntary: Firm initiated · Other
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Voluntis
- Distributor
- Distributed to accounts in Arizona, Masachusetts, Michigan, Ohio, Tennessee, and Vermont. International distribution to Canada and France (however, French customers are not affected by this recall).
- Lot
- Software version numbers 1.4.3 / 1.6.3 / 1.6.31 / 1.6.40
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- France
Related recalls
- FDAINFed (Iron Dextran Injection USP) 100 mg elemental iron/2 mL (50 mg/mL), Rx only, packaged in 2 mL SterileDecember 5, 2018
- FDARecchiuti Bittersweet, Smooth, rich and complex. Net Wt. 3 oz. (85 G) Also sold in case of 12 UPC 8019895November 28, 2018
- FDAPremierPro Aluminum Adjustable WalkerNovember 28, 2018
- FDAHOUSE BLEND Barbecue SAUCE INGREDIENTS: ***Contains 2% or less of Soybean Oil***Onion Powder***CalciumNovember 28, 2018
- FDAVanguard Knee System -PS Open Box Femoral Left, 62.5 mm Item Number: 183126November 28, 2018
- FDASquare Shell for Turnovers, Elaboracion Felo, Inc. Avenida Yumet No. 112, Aguadilla, PR 00603November 28, 2018
Questions about this recall
Was Insulia Diabetes Management Companion recalled?
Yes. FDA published recall Z-0478-2019 on November 28, 2018. Reported hazard: The firm identified an issue with the basal calculator identified on the Android version of the Insulia application. The bug could lead from low-impact to high-impact hypoglycemia depending on the circumstances. A new update of Insulia is now available (version 1.6.42), which resolves this issue.
What should I do if I have this product?
Voluntary: Firm initiated · Other