Insulia Diabetes Management Companion

Published November 28, 2018 · Recall initiated: October 8, 2018

Summary

FDA records medical device Class II recall Z-0478-2019, published November 28, 2018 and initiated October 8, 2018. Recalling firm: Voluntis; Status: Terminated; Reported quantity: 14; Lot: Software version numbers 1.4.3 / 1.6.3 / 1.6.31 / 1.6.40. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The firm identified an issue with the basal calculator identified on the Android version of the Insulia application. The bug could lead from low-impact to high-impact hypoglycemia depending on the circumstances. A new update of Insulia is now available (version 1.6.42), which resolves this issue.

What to do

Voluntary: Firm initiated · Other

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Voluntis
Distributor
Distributed to accounts in Arizona, Masachusetts, Michigan, Ohio, Tennessee, and Vermont. International distribution to Canada and France (however, French customers are not affected by this recall).
Lot
Software version numbers 1.4.3 / 1.6.3 / 1.6.31 / 1.6.40
Risk classification
Class II
Area (FDA)
device
Country
France

Related recalls

Questions about this recall

Was Insulia Diabetes Management Companion recalled?

Yes. FDA published recall Z-0478-2019 on November 28, 2018. Reported hazard: The firm identified an issue with the basal calculator identified on the Android version of the Insulia application. The bug could lead from low-impact to high-impact hypoglycemia depending on the circumstances. A new update of Insulia is now available (version 1.6.42), which resolves this issue.