Allegedly counterfeit batteries manufactured by CSB batteries being used with Plum Infusion Systems: Plum 360" Infusion System, Pump List Number 30010; Plum A+ & Plum A+3" Infusion Systems, Pump List Numbers: 11005, 11971, 12391, 12618, 20678, 20679, 20792, 60529, 12348, 1197
Published November 27, 2024 · Recall initiated: October 22, 2024
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
ICU Medical has received reports of allegedly counterfeit CSB batteries being used with Plum Infusion Systems. While these non-OEM batteries are visually similar to the Plum batteries supplied by ICU Medical, they are in fact not the same batteries and have not been tested or validated for use with Plum Infusion Systems. Preliminary reports suggest that these non-OEM batteries fail to hold their charge and the pump may display messages to replace batteries earlier than expected. OSI Batteries and their customers are distributing these non-OEM CSB batteries without authorization from ICU Medical.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- ICU Medical, Inc.
- Distributor
- Worldwide distribution: US (nationwide) including Puerto Rico and OUS (foreign) countries of: Argentina, Brazil, Canada, Chile, Colombia, France, Greece, Hong Kong, Italy, Ireland Jordan, Kuwait, Mexico, Netherlands, Oman, Peru, Philippines, Poland, Saudia Arabia, Spain, Taiwan, Turkey, United Kingdom, and Uruguay.
- Lot
- All batteries WITHOUT the following information on the label: 1) ICU Medlcal Test Label, 2) CE Mark, and WITH a Date code (yellow label found on side of battery): W2401xxxx - W2406xxx
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States