FreeStyle Lite Blood Glucose Test Strips; For in vitro diagnostic testing. 50 count; 100 count Product of Ireland; UPC 6 99073 70819 9; - 50 count UPC 6 99073 70822 9 - 50 count UPC 6 99073 71026 0 - 50 count UPC 6 99073 70827 4 - 100 count Manufactured by Abbott Diabetes Care Inc. Alameda, CA. The FreeStyle Blood Glucose test strip is intended for use in the quantitative measurement of glucose in capillary whole blood from the finger, upper arm and palm. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of a diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and it is not intended for use on neonates or arterial blood.
Published December 25, 2013 · Recall initiated: November 18, 2013
Summary
FDA records medical device Class I recall Z-0486-2014, published December 25, 2013 and initiated November 18, 2013. Recalling firm: Abbott Diabetes Care, Inc.; Status: Terminated; Reported quantity: 141,219 cartons for human use - total; Lot: Part number: 71026-70; Lot number: 1281732; Exp. date: 2014/05; Lot number: 1283603; Exp. date: 2014/06; Part number: 70827-71: Lot number: 1363015, Exp. Date: 2014/11; Lot number: 1363109, Exp. Date 2014/11: Lot number: 1365921, Exp. Date 2014/12; Part number 70827-72: Lot number: 1366006 Exp. Date : 2014/12; Lot number 1366111, Exp. Date: 2014/12; Lot number 1366337; Exp. Date: 2014/12; Lot number: 1366515, Exp Date: 2014/12; Part number 70822-71; Lot number 1283345: Exp. Date: 2014/05; Lot number 1365920; Exp. Date: 2014/12; Lot number 1365934; Exp. Date: 2014/12; Part number: 70822-72: Lot number: 1350414: Exp Date: 2014/07; Lot number 1363321; Exp. Date 2014/11: Part number: 70819-70: Lot Number: 1365056: Exp. Date 2014/11; Lot number: 1366347: Exp. Date: 2014/12; Lot number: 1367917; Exp. Date 2014/12 Lot number: 1374907; Exp. Date 2015/03. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Certain lots of FreeStyle and FreeStyle Lite Blood Glucose Test Strips produce erroneously low blood glucose results when using FreeStyle Blood Glucose Meters, FreeStyle Flash Blood Glucose Meters and the FreeStyle blood glucose meter built into the OmniPod system. Erroneously low results that are not recognized may pose significant risks to your health.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Abbott Diabetes Care, Inc.
- Distributor
- Worldwide Distribution - USA (nationwide and Puerto Rico) and Internationally to Australia, Austria, Belgium, Denmark, Egypt, Finland, France, Germany, Jamaica, Ireland, Israel, United Kingdom, Luxembourg, Netherlands, Norway, Sweden, Switzerland, UAE, and Canada. **Center Recommended Depth - Consumers/User**
- Lot
- Part number: 71026-70; Lot number: 1281732; Exp. date: 2014/05; Lot number: 1283603; Exp. date: 2014/06; Part number: 70827-71: Lot number: 1363015, Exp. Date: 2014/11; Lot number: 1363109, Exp. Date 2014/11: Lot number: 1365921, Exp. Date 2014/12; Part number 70827-72: Lot number: 1366006 Exp. Date : 2014/12; Lot number 1366111, Exp. Date: 2014/12; Lot number 1366337; Exp. Date: 2014/12; Lot number: 1366515, Exp Date: 2014/12; Part number 70822-71; Lot number 1283345: Exp. Date: 2014/05; Lot number 1365920; Exp. Date: 2014/12; Lot number 1365934; Exp. Date: 2014/12; Part number: 70822-72: Lot number: 1350414: Exp Date: 2014/07; Lot number 1363321; Exp. Date 2014/11: Part number: 70819-70: Lot Number: 1365056: Exp. Date 2014/11; Lot number: 1366347: Exp. Date: 2014/12; Lot number: 1367917; Exp. Date 2014/12 Lot number: 1374907; Exp. Date 2015/03.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAFreeStyle Libre 3 Sensors, REF: 72081-01, component of the FreeStyle Libre 3 Continuous Glucose MonitoringSeptember 4, 2024
- FDAFreeStyle Libre 3 App, part of Continuous Glucose Monitoring System, when used on the Android 13 OperatingJanuary 24, 2024
- FDAPrecision Xtra Blood Glucose & Ketone Monitoring System, UDI: 00093815715023, 00093815988144April 1, 2020
- FDAFreeStyle Blood Glucose Monitors System. For in vitro diagnostic use only. Store at room temperature below 86March 26, 2014
- FDAFreeStyle Blood Glucose Test Strips; For in vitro diagnostic testing. 50 count UPC 6 99073 12050 2; FreestyleDecember 25, 2013
- FDANephros SafeSpout Manufacturer: Nephros Inc. 41 Grand Avenue, River Edge, NJ 07661 Assembled at : MEDICA 41036December 25, 2013
Questions about this recall
Was FreeStyle Lite Blood Glucose Test Strips; For in vitro diagnostic testing. 50 count; 100 count Product of Ireland; UPC 6 99073 70819 9; - 50 count UPC 6 99073 70822 9 - 50 count UPC 6 99073 71026 0 - 50 count UPC 6 99073 70827 4 - 100 count Manufactured by Abbott Diabetes Care Inc. Alameda, CA. The FreeStyle Blood Glucose test strip is intended for use in the quantitative measurement of glucose in capillary whole blood from the finger, upper arm and palm. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of a diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and it is not intended for use on neonates or arterial blood. recalled?
Yes. FDA published recall Z-0486-2014 on December 25, 2013. Reported hazard: Certain lots of FreeStyle and FreeStyle Lite Blood Glucose Test Strips produce erroneously low blood glucose results when using FreeStyle Blood Glucose Meters, FreeStyle Flash Blood Glucose Meters and the FreeStyle blood glucose meter built into the OmniPod system. Erroneously low results that are not recognized may pose significant risks to your health.
What should I do if I have this product?
Voluntary: Firm initiated · Letter