Syngo.plaza Picture Archiving and Communication System (PACS) with software version VB20A; Model numbers: 10863171, 10863172, 10863173 Syngo.plaza is a Picture Archiving and Communication System (PACS) software device intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. Syngo.plaza also supports storage and archiving of DICOM Structured reports. In a comprehensive imaging syngo.plaza integrates Hospital/Radiology Information Systems (HIS/RIS) to enable customer specific workflows. Syngo.plaza optionally uses a variety of advanced postprocessing applications.

Published February 7, 2018 · Recall initiated: August 18, 2017

Summary

FDA records medical device Class II recall Z-0495-2018, published February 7, 2018 and initiated August 18, 2017. Recalling firm: Siemens Medical Solutions USA, Inc; Status: Terminated; Reported quantity: 47; Lot: Syngo.plaza with software version VB20A; Model numbers: 10863171, 10863172, 10863173. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Software update to correct several issues that include (1)Potential data loss, (2) Study mix-up, (3) Incorrect measurements on multi-frame images, (4) Dearchiving issue, and (5) Unauthorized access of data due to inadequate permissions for shared folders.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Siemens Medical Solutions USA, Inc
Distributor
USA (nationwide) Distribution to the states of : AL, CT, FL, KS, LA, MA, MI, MN, MO, MS, NE, NY, OH, OR, PA, TX and WI. Foreign consignees were not reported as device is manufactured outside the U.S.
Lot
Syngo.plaza with software version VB20A; Model numbers: 10863171, 10863172, 10863173
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Syngo.plaza Picture Archiving and Communication System (PACS) with software version VB20A; Model numbers: 10863171, 10863172, 10863173 Syngo.plaza is a Picture Archiving and Communication System (PACS) software device intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. Syngo.plaza also supports storage and archiving of DICOM Structured reports. In a comprehensive imaging syngo.plaza integrates Hospital/Radiology Information Systems (HIS/RIS) to enable customer specific workflows. Syngo.plaza optionally uses a variety of advanced postprocessing applications. recalled?

Yes. FDA published recall Z-0495-2018 on February 7, 2018. Reported hazard: Software update to correct several issues that include (1)Potential data loss, (2) Study mix-up, (3) Incorrect measurements on multi-frame images, (4) Dearchiving issue, and (5) Unauthorized access of data due to inadequate permissions for shared folders.

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail