Abbott Navitor Transcatheter Aortic Heart Valve, 27 mm, NVTR-27, Sterile
Published November 27, 2024 · Recall initiated: October 17, 2024
Summary
FDA records medical device Class II recall Z-0495-2025, published November 27, 2024 and initiated October 17, 2024. Recalling firm: Abbott Medical; Status: Ongoing; Reported quantity: 5 OUS; Lot: UDI-DI (GTIN): 5415067031594, Serial Numbers: 20110016, 20155472, 20155476, 20166304, & 20168050. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A manufacturing error resulting in acceptance of valves with a leaflet deflection value that was outside of established specifications. This could potentially affect long-term durability of the impacted valve and/or cause potential valve failure.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Abbott Medical
- Distributor
- New York, Australia, Belgium, Germany, Iran, Israel, Japan, Kazakhstan, Oman, Poland, Romania, Slovakia, Spain, Turkey, United Kingdom.
- Lot
- UDI-DI (GTIN): 5415067031594, Serial Numbers: 20110016, 20155472, 20155476, 20166304, & 20168050.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMerlin" Patient Care System (PCS) software Merlin" PCS 3650 programmer Model 3330 softwareJune 17, 2026
- FDAMerlin@home with the MerlinOnDemand capability Transmitter REF EX1150 Merlin@home with the MerlinOnDemand"December 4, 2024
- FDAMerlin@home Transmitter REF EX1100 The Merlin@home" (M@h) transmitter is intended to transfer stored data fromDecember 4, 2024
- FDAAbbott Navitor, Vision Technology, Transcatheter Aortic Heart Valve, 27 mm, NVRO-27, SterileNovember 27, 2024
- FDASiemens Atellica CH Reaction Ring Cuvette Segment-has cuvettes that contain and incubate the sample reactionNovember 27, 2024
- FDAAbbot Navitor, Vision Technology, Transcatheter Aortic Heart Valve, 29 mm, NVRO-29, SterileNovember 27, 2024
Questions about this recall
Was Abbott Navitor Transcatheter Aortic Heart Valve, 27 mm, NVTR-27, Sterile recalled?
Yes. FDA published recall Z-0495-2025 on November 27, 2024. Reported hazard: A manufacturing error resulting in acceptance of valves with a leaflet deflection value that was outside of established specifications. This could potentially affect long-term durability of the impacted valve and/or cause potential valve failure.
What should I do if I have this product?
Voluntary: Firm initiated · Letter