Defibtech DDP-2002: Pediatric Defibrillation Pads For use with AED Models: Lifeline View Auto, lifeline Pro, Lifeline ECG, Lifeline View (DDU-2XXX Series -Indicated for use on victims of sudden cardiac arrest (SCA.)
Published December 4, 2024 · Recall initiated: October 14, 2024
Summary
FDA records medical device Class II recall Z-0505-2025, published December 4, 2024 and initiated October 14, 2024. Recalling firm: Defibtech, LLC; Status: Ongoing; Reported quantity: 22,294 total; Lot: UDI-DI: 00815098020454, DDP-2002, pediatric defibrillation pads, single pair; 10815098020451, DDP-2002, pediatric defibrillation pads, single pair configuration; 40815098020452, DDP-2002, pediatric defibrillation pads, 20 pairs. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Within the Irish Market, that a Field Safety Notice is being issued due to a select number of authorized labelled Defibtech AED Pads, which have had the authorized label removed and an unauthorized label applied, could potentially lead to expired Defibtech AED Pads appearing as though they are within the acceptable shelf life.there is a possibility of hydrogel drying, which may result in poor electrical connection to the patient, which can reduce the energy delivered to the patient
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Defibtech, LLC
- Distributor
- Ireland
- Lot
- UDI-DI: 00815098020454, DDP-2002, pediatric defibrillation pads, single pair; 10815098020451, DDP-2002, pediatric defibrillation pads, single pair configuration; 40815098020452, DDP-2002, pediatric defibrillation pads, 20 pairs
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDADefibtech DDP-2001: Adult Defibrillation Pads For use with AED Models: Lifeline View Auto, lifeline Pro,December 4, 2024
- FDADefibtech DDP-200P: Pediatric Defibrillation Pads 0 For use with AED Models: Lifeline and Lifeline Auto (DDU-lDecember 4, 2024
- FDADefibtech DDP-100 Adult Defibrillation Pads For use with AED Models: Lifeline and Lifeline Auto (DDU-1:XXDecember 4, 2024
- FDAV.A.C. Ulta Therapy Unit REF CAPULTDEV01 & ULTDEV01 The 3M" V.A.C.¿ Ulta Negative Pressure Wound TherapyDecember 4, 2024
- FDAMEDLINE NEXUS 19/20G CATH CONNECTOR, Medline SKU REF NEXCONNECT, Catheter ConnectorDecember 4, 2024
- FDARMU-2000 Automated Chest Compression DeviceAugust 21, 2024
Questions about this recall
Was Defibtech DDP-2002: Pediatric Defibrillation Pads For use with AED Models: Lifeline View Auto, lifeline Pro, Lifeline ECG, Lifeline View (DDU-2XXX Series -Indicated for use on victims of sudden cardiac arrest (SCA.) recalled?
Yes. FDA published recall Z-0505-2025 on December 4, 2024. Reported hazard: Within the Irish Market, that a Field Safety Notice is being issued due to a select number of authorized labelled Defibtech AED Pads, which have had the authorized label removed and an unauthorized label applied, could potentially lead to expired Defibtech AED Pads appearing as though they are within the acceptable shelf life.there is a possibility of hydrogel drying, which may result in poor electrical connection to the patient, which can reduce the energy delivered to the patient
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail