SonarMed AirWave Monitor Kit (Monitor, Sensor Cable, Power Cord), Item Code AW-M0001. used with endotracheal tube (ETT)
Published January 26, 2022 · Recall initiated: December 22, 2021
Summary
FDA records medical device Class II recall Z-0514-2022, published January 26, 2022 and initiated December 22, 2021. Recalling firm: Covidien, LP; Status: Ongoing; Reported quantity: 66 devices; Lot: UDI-Device Identifier (GTIN/UPN): 00851334007001; Serial Numbers: AW0501 AW0519 AW0537 AW0556 AW0502 AW0520 AW0540 AW0557 AW0503 AW0521 AW0541 AW0558 AW0504 AW0522 AW0542 AW0559 AW0505 AW0524 AW0543 AW0560 AW0506 AW0525 AW0544 AW0561 AW0507 AW0526 AW0545 AW0562 AW0508 AW0527 AW0546 AW0563 AW0509 AW0528 AW0548 AW0564 AW0510 AW0529 AW0549 AW0565 AW0511 AW0530 AW0550 AW0566 AW0512 AW0531 AW0551 AW0567 AW0513 AW0532 AW0552 AW0568 AW0514 AW0533 AW0553 AW0569 AW0515 AW0536 AW0554 AW0570 AW0516 AW0538 AW0555 AW0571 AW0517 AW0539. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Prior to use on a patient, the device alarm speaker failed to annunciate on power up and the device falsely detected a sensor cable disconnect and reconnect during setup. The monitor detection of a false sensor cable disconnect and reconnect may cause the monitor to discontinue monitoring and transition into the calibration screen without an audible alarm.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Covidien, LP
- Distributor
- The devices were distributed to the following US states: CA, CO, FL, IL, KY, MA, OH, TN, and TX.
- Lot
- UDI-Device Identifier (GTIN/UPN): 00851334007001; Serial Numbers: AW0501 AW0519 AW0537 AW0556 AW0502 AW0520 AW0540 AW0557 AW0503 AW0521 AW0541 AW0558 AW0504 AW0522 AW0542 AW0559 AW0505 AW0524 AW0543 AW0560 AW0506 AW0525 AW0544 AW0561 AW0507 AW0526 AW0545 AW0562 AW0508 AW0527 AW0546 AW0563 AW0509 AW0528 AW0548 AW0564 AW0510 AW0529 AW0549 AW0565 AW0511 AW0530 AW0550 AW0566 AW0512 AW0531 AW0551 AW0567 AW0513 AW0532 AW0552 AW0568 AW0514 AW0533 AW0553 AW0569 AW0515 AW0536 AW0554 AW0570 AW0516 AW0538 AW0555 AW0571 AW0517 AW0539
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAV-Loc 180 Absorbable Wound Closure Device BOX BEDSIDE STAPLING IS KING BOX06614V4 BOX BY PASS FINAL BOX07405V1August 19, 2026
- FDASurgipro II S8698GX SURGIPRO II BL 5/0 18 P-13 S8698GXAugust 19, 2026
- FDAV-Loc 90 Absorbable Wound Closure Device VLOC 90 0 VL 22 GS-21 VLOCM0336 VLOC 90 0 VL 22 GS-25 VLOCM0436August 19, 2026
- FDADermalon 88861757-31 DERMALON* 4-0 BLU 75CM C13 88861757-31 88861799-31 DERMALON* 4-0 BLU 75CM C14 88861799-31August 19, 2026
- FDAMetformin Hydrochloride Extended-Release Tablets, USP 750 mg, 100 count HDPE bottles NDC# 72578-036-01February 2, 2022
- FDAInfa-Therm Transport Mattress, Reference Number 989805616831 1015January 26, 2022
Questions about this recall
Was SonarMed AirWave Monitor Kit (Monitor, Sensor Cable, Power Cord), Item Code AW-M0001. used with endotracheal tube (ETT) recalled?
Yes. FDA published recall Z-0514-2022 on January 26, 2022. Reported hazard: Prior to use on a patient, the device alarm speaker failed to annunciate on power up and the device falsely detected a sensor cable disconnect and reconnect during setup. The monitor detection of a false sensor cable disconnect and reconnect may cause the monitor to discontinue monitoring and transition into the calibration screen without an audible alarm.
What should I do if I have this product?
Voluntary: Firm initiated · Letter