Back to search
FDAFDA · Food and drugs·Recall number Z-0514-2026

MEDLINE convenience kits labeled as: 1) CENTRAL LINE PACK, REF DYNJ00281P; 2) FREE FLAP PACK, REF DYNJ68213C.

Published November 26, 2025 · Recall initiated: September 30, 2025

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Recall details

Manufacturer
Medline Industries, LP
Distributor
US Nationwide distribution.
Lot
1) REF DYNJ00281P: UDI/DI 10195327411954 (each), 40195327411955 (case), Lot Numbers: 23GMD333, 23HMC426; 2) REF DYNJ68213C: UDI/DI 10195327511999 (each), 40195327511990 (case), Lot Numbers: 23IMI021, 23KMJ671.
Risk classification
Class II
Area (FDA)
device
Country
United States