Total Temporomandibular Joint Replacement System- 50mm Rt Narrow Ti Mand Reconstruction of the temporomandibular joint

Published December 17, 2014 · Recall initiated: October 17, 2014

Summary

FDA records medical device Class II recall Z-0515-2015, published December 17, 2014 and initiated October 17, 2014. Recalling firm: Biomet Microfixation, LLC; Status: Terminated; Reported quantity: 42; Lot: Model 01-6550TI, Lot #s: 531890A, 531890B, 531900A, 531900B, 544220B, 544230A, 544230B, 544200A, 544200B, 544210A, 544210B, 544220A. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Laser etching on the parts is wider and deeper than the conditions previously validated.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Biomet Microfixation, LLC
Distributor
Worldwide Distribution - USA including PA, MN, CA and Internationally to Australia, Netherlands, France, and United Kingdom.
Lot
Model 01-6550TI, Lot #s: 531890A, 531890B, 531900A, 531900B, 544220B, 544230A, 544230B, 544200A, 544200B, 544210A, 544210B, 544220A.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Total Temporomandibular Joint Replacement System- 50mm Rt Narrow Ti Mand Reconstruction of the temporomandibular joint recalled?

Yes. FDA published recall Z-0515-2015 on December 17, 2014. Reported hazard: Laser etching on the parts is wider and deeper than the conditions previously validated.

What should I do if I have this product?

Voluntary: Firm initiated · Letter