Kreatech PapV-16, Flu labeled ISH probe, to detect presence of HPV-16 DNA inserts in human tissue, Model Number 40V016V495
Published December 4, 2024 · Recall initiated: November 1, 2024
Summary
FDA records medical device Class II recall Z-0522-2025, published December 4, 2024 and initiated November 1, 2024. Recalling firm: Kreatech Biotechnology Bv; Status: Ongoing; Reported quantity: 23 units; Lot: UDI-DI: (01)08718858480241 Lot Number: 00088454. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The DNA fragments for this specific lot number are too large, resulting in potential background around the cells due to suboptimal hybridization. An absence of signal could also be observed as a result of the suboptimal hybridization of the larger DNA fragments. The user may interpret potential background as a false positive result and an absence of signal as a false negative result, if proper positive and negative controls are not included.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Kreatech Biotechnology Bv
- Distributor
- US States: FL, NY
- Lot
- UDI-DI: (01)08718858480241 Lot Number: 00088454
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Netherlands
Related recalls
- FDABrand Name: MST Duet Reusable Handle for Single Use Heads Product Name: Reusable Handle for Single Use HeadsDecember 4, 2024
- FDABrand Name: Z-800 Infusion System Product Name: Z-800WF Model/Catalog Number: Z-800WF. The Z-800 InfusionDecember 4, 2024
- FDATomographic Imager Combining Emission Computed Tomography With Nuclear Magnetic Resonance: 1) SIGNA PET/MR, 2)December 4, 2024
- FDA4.5 x 180 mm REF 28208BKS Aggressive Cutter, sterile, 6x 3.5 x 70 mm REF 28206CBS Full Radius Resector,December 4, 2024
- FDASIGNA Architect AIR, Nuclear Magnetic Resonance Imaging SystemDecember 4, 2024
- FDAsmiths medical portex Venn Reusable Tracheal Tube Introducer, 15Ch x 60cm, REF 14-504-17 and REF 14-504-17JPDecember 4, 2024
Questions about this recall
Was Kreatech PapV-16, Flu labeled ISH probe, to detect presence of HPV-16 DNA inserts in human tissue, Model Number 40V016V495 recalled?
Yes. FDA published recall Z-0522-2025 on December 4, 2024. Reported hazard: The DNA fragments for this specific lot number are too large, resulting in potential background around the cells due to suboptimal hybridization. An absence of signal could also be observed as a result of the suboptimal hybridization of the larger DNA fragments. The user may interpret potential background as a false positive result and an absence of signal as a false negative result, if proper positive and negative controls are not included.
What should I do if I have this product?
Voluntary: Firm initiated · Letter