Praxair Vantage Grab n Go - Gas Cylinder pressure regulator Catalog Number: OX-M-AEGNGVNTG Usage: Regulator reduces pressure in oxygen-containing cylinder.
Published January 2, 2013 · Recall initiated: November 16, 2012
Summary
FDA records medical device Class I recall Z-0525-2013, published January 2, 2013 and initiated November 16, 2012. Recalling firm: Praxair Inc.; Status: Terminated; Reported quantity: 320,451 units (255,690 US ; 55,884 Canada; 8877 OUS); Lot: OX M-AEGNGVNTG (US and Canada) and WESPRX-9500 (Global) Cyclinders in service since Jan 2009. Units with a "T" decal (T-teflon) on the fill side of the valve have the rework completed. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Isolated incidents of ignition inside Grab n' Go cylinders that had been knocked over or otherwise subject to significant physical impact.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Praxair Inc.
- Distributor
- Worldwide distribution: USA (nationwide) and countries of: Brazil, Canada, China, Columbia, Costa Rica, India, Paraguay and Uruguay.
- Lot
- OX M-AEGNGVNTG (US and Canada) and WESPRX-9500 (Global) Cyclinders in service since Jan 2009. Units with a "T" decal (T-teflon) on the fill side of the valve have the rework completed.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAUN1002 Air, Compressed, Medical Air, USP, Distributed by Praxair Distribution, Inc. Roseville, MN 55113,April 6, 2016
- FDAUN1002 Air, Compressed, Breathing Grade, USP, Distributed by Praxair Distribution, Inc. Roseville, MN 55113,April 6, 2016
- FDAComputed Tomography X-Ray System, Extended Brilliance Workspace (EBW), Model #728260 Philips Healthcare,January 2, 2013
- FDAZimmer Spine Ardis Inserter, Rx, non-sterileJanuary 2, 2013
- FDATaylor Farms Hearts of Romaine Lettuce, All Natural, Fresh Hand Selected, Triple Washed and Ready to Enjoy. 10January 2, 2013
- FDAGE Giraffe and Panda T-Piece Resuscitation System labeled ***T-Piece Neonatal Patient Circuit Kit DisposableJanuary 2, 2013
Questions about this recall
Was Praxair Vantage Grab n Go - Gas Cylinder pressure regulator Catalog Number: OX-M-AEGNGVNTG Usage: Regulator reduces pressure in oxygen-containing cylinder. recalled?
Yes. FDA published recall Z-0525-2013 on January 2, 2013. Reported hazard: Isolated incidents of ignition inside Grab n' Go cylinders that had been knocked over or otherwise subject to significant physical impact.
What should I do if I have this product?
Voluntary: Firm initiated · Letter