Distal Access Catheter: Model Numbers: 90121, 90130, 90131, 90160. The Distal Access Catheter is intended for use in facilitating the insertion and guidance of an occlusion catheter, infusion catheter or other appropriate microcatheter into a selected blood vessel in the peripheral, coronary and neuro vascular systems.

Published December 17, 2014 · Recall initiated: September 26, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Distal Access Catheters were shipped to US customers with the Directions for Use that included instructions for aspiration. This version of the DFU is only approved outside of the US.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Concentric Medical Inc
Distributor
US Distribution to states of: TX, TN, PA, MI, VA, CA, NC, MS, FL, IL, MA, AZ, OK, NY, KY, WA, and AL.
Lot
Lot Number 36007: 09-2014 Lot Number 35767: 05-2014 Lot Number 35980: 08-2014 Lot Number 35951: 08-2014 Lot Number 36050: 09-2014 Lot Number 36017: 09-2014 Lot Number 36029: 09-2014 Lot Number 35583: 02-2014
Risk classification
Class II
Area (FDA)
device
Country
United States