Back to search
FDAFDA · Food and drugs·Recall number Z-0526-2015

Varian On-Board Imager Advanced Imaging System. The On-Board Imager device is used for verification of correct patient position in relation to isocenter and verification of the treatment fields in relation to anatomical and/or fiducial land®¿ marks.

Published December 24, 2014 · Recall initiated: April 30, 2014

Summary

FDA records medical device Class II recall Z-0526-2015, published December 24, 2014 and initiated April 30, 2014. Recalling firm: Varian Medical Systems, Inc.; Status: Terminated; Reported quantity: 32 units; Lot: Part Number P1003290001. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Varian MedicalSystems' X-ray generator Supplier (CPI) has identified that 32 VMS150 X-ray Generators were shipped that are missing a ground wire between the EMIbraid on the induc­ tion assembly cable and the ground stud on the lid of the High Voltage Module.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Varian Medical Systems, Inc.
Distributor
Worldwide Distribution.
Lot
Part Number P1003290001
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Varian On-Board Imager Advanced Imaging System. The On-Board Imager device is used for verification of correct patient position in relation to isocenter and verification of the treatment fields in relation to anatomical and/or fiducial land®¿ marks. recalled?

Yes. FDA published recall Z-0526-2015 on December 24, 2014. Reported hazard: Varian MedicalSystems' X-ray generator Supplier (CPI) has identified that 32 VMS150 X-ray Generators were shipped that are missing a ground wire between the EMIbraid on the induc­ tion assembly cable and the ground stud on the lid of the High Voltage Module.

What should I do if I have this product?

Voluntary: Firm initiated · Letter