Varian On-Board Imager Advanced Imaging System. The On-Board Imager device is used for verification of correct patient position in relation to isocenter and verification of the treatment fields in relation to anatomical and/or fiducial land®¿ marks.
Published December 24, 2014 · Recall initiated: April 30, 2014
Summary
FDA records medical device Class II recall Z-0526-2015, published December 24, 2014 and initiated April 30, 2014. Recalling firm: Varian Medical Systems, Inc.; Status: Terminated; Reported quantity: 32 units; Lot: Part Number P1003290001. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Varian MedicalSystems' X-ray generator Supplier (CPI) has identified that 32 VMS150 X-ray Generators were shipped that are missing a ground wire between the EMIbraid on the induc­ tion assembly cable and the ground stud on the lid of the High Voltage Module.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Varian Medical Systems, Inc.
- Distributor
- Worldwide Distribution.
- Lot
- Part Number P1003290001
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAVarian Head Frame - Model number HHF -ALL Head Frame Posts (PN1008016 ) and Post Kits (PN1008016). The VarianOctober 19, 2016
- FDAOptical Surface Monitoring System{OSMS, Varian Private Label} This is used on Varian's TrueBeam EDGE;July 20, 2016
- FDAIntended for both adult and pediatric patients for taking diagnostic radiographic exposures of the skull,December 24, 2014
- FDAPKG, BIPOLAR FORCEPS, MICRO TIP, 45CM, P/N 0250080115 nationwide in US and worldwide: Algeria, Argentina,December 24, 2014
- FDAC-Series High Energy Linear Accelerator - Models: Novalis Tx, Trilogy, Trilogy Tx, Clinac iX, Clinac 2100 C/D,September 24, 2014
- FDA4D Integrated Treatment Console (4DITC), versions 10 and 11. 4DITC provided assistance for accurate treatmentSeptember 24, 2014
Questions about this recall
Was Varian On-Board Imager Advanced Imaging System. The On-Board Imager device is used for verification of correct patient position in relation to isocenter and verification of the treatment fields in relation to anatomical and/or fiducial land®¿ marks. recalled?
Yes. FDA published recall Z-0526-2015 on December 24, 2014. Reported hazard: Varian MedicalSystems' X-ray generator Supplier (CPI) has identified that 32 VMS150 X-ray Generators were shipped that are missing a ground wire between the EMIbraid on the induc­ tion assembly cable and the ground stud on the lid of the High Voltage Module.
What should I do if I have this product?
Voluntary: Firm initiated · Letter