Varian On-Board Imager Advanced Imaging System. The On-Board Imager device is used for verification of correct patient position in relation to isocenter and verification of the treatment fields in relation to anatomical and/or fiducial land¿¿ marks.

Published December 24, 2014 · Recall initiated: April 30, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Varian MedicalSystems' X-ray generator Supplier (CPI) has identified that 32 VMS150 X-ray Generators were shipped that are missing a ground wire between the EMIbraid on the induc­ tion assembly cable and the ground stud on the lid of the High Voltage Module.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Varian Medical Systems, Inc.
Distributor
Worldwide Distribution.
Lot
Part Number P1003290001
Risk classification
Class II
Area (FDA)
device
Country
United States