MEDLINE convenience kits labeled as: GP-LEEP PACK-LF, REF PHS396969004F
Published November 26, 2025 · Recall initiated: September 30, 2025
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Recall details
- Manufacturer
- Medline Industries, LP
- Distributor
- US Nationwide distribution.
- Lot
- UDI/DI 10195327454227 (each) 40195327454228 (case), Lot Numbers: 23HDA742
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States