Zimmer M/DN® Intramedullary Fixation Humeral Guide Wire - Bullet Tip, 2.4 mm Diameter, 70 cm Length Item Number: 00-2255-026-00 Product Usage: A single use surgical instrument used during orthopaedic surgery for implantation of a prosthesis.
Published December 12, 2018 · Recall initiated: October 22, 2018
Summary
FDA records medical device Class II recall Z-0532-2019, published December 12, 2018 and initiated October 22, 2018. Recalling firm: Zimmer Biomet, Inc.; Status: Terminated; Reported quantity: 780; Lot: All lots expiring prior to September 30, 2023. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Various Trauma Guide Wires- creases in the sealing area on the end opposite of the product's chevron opening compromising product sterility
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Zimmer Biomet, Inc.
- Distributor
- Worldwide - US Nationwide Distribution Foreign: CANADA AUSTRALIA BRAZIL CHILE HONG KONG JAPAN MALAYSIA NETHERLANDS NICARAGUA SINGAPORE TAIWAN
- Lot
- All lots expiring prior to September 30, 2023
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASpinal Rod Cutter, Item number 00392500200. The product is a nonsterile pin cutter - Product Usage: TheDecember 30, 2020
- FDAZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Snap-In O.D. Cup with Spacers, 55November 11, 2020
- FDAZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Neutral, O.D. Cup with Spacers, 55November 11, 2020
- FDAZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, 10 degree inclined face, O.D. CupNovember 11, 2020
- FDAGel Alkaline Phosphatase Isoenzyme Control Kit: Gel Alkaline Phosphatase Isoenzyme Control Cat. No. 5104 1 xDecember 12, 2018
- FDAZimmer ZMS® Intramedullary Fixation Smooth Guide Wire, 2.4 mm Diameter, 100 cm Length Item Number:December 12, 2018
Questions about this recall
Was Zimmer M/DN® Intramedullary Fixation Humeral Guide Wire - Bullet Tip, 2.4 mm Diameter, 70 cm Length Item Number: 00-2255-026-00 Product Usage: A single use surgical instrument used during orthopaedic surgery for implantation of a prosthesis. recalled?
Yes. FDA published recall Z-0532-2019 on December 12, 2018. Reported hazard: Various Trauma Guide Wires- creases in the sealing area on the end opposite of the product's chevron opening compromising product sterility
What should I do if I have this product?
Voluntary: Firm initiated · Letter