Back to search
FDAFDA · Food and drugs·Recall number Z-0536-2013

Biomet Polaris 5.5 Button Lock Screw Inserter. Catalog number 14-500178

Published December 26, 2012 · Recall initiated: September 29, 2012

Summary

FDA records medical device Class II recall Z-0536-2013, published December 26, 2012 and initiated September 29, 2012. Recalling firm: Ebi, Llc; Status: Terminated; Reported quantity: 15 units; Lot: Lot Numbers 540199, 540199-1. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Broken screw insertor hardness was measured and found to be below the acceptable limits.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Ebi, Llc
Distributor
Nationwide Distribution, including the states of MO, TX, and IN.
Lot
Lot Numbers 540199, 540199-1
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Biomet Polaris 5.5 Button Lock Screw Inserter. Catalog number 14-500178 recalled?

Yes. FDA published recall Z-0536-2013 on December 26, 2012. Reported hazard: Broken screw insertor hardness was measured and found to be below the acceptable limits.

What should I do if I have this product?

Voluntary: Firm initiated · Letter