MagNA Pure 96 System Version 1.0 Software Version 2.0 Robotic workstation for automated isolation of nucleic acids from different kinds of crude sample material (whole blood, serum, blood cells, culture cells, tissue, bacteria, fungi) by the use of the specially designed MagNA Pure reagent kits, automated filling of different kinds of PCR reaction vessels.
Published December 25, 2013 · Recall initiated: November 25, 2013
Summary
FDA records medical device Class II recall Z-0536-2014, published December 25, 2013 and initiated November 25, 2013. Recalling firm: Roche Diagnostics Operations, Inc.; Status: Terminated; Reported quantity: 27; Lot: Serial Numbers 2001, 2002, 2003, 2004, 2006, 2011, 2012, 2013, 2014, 2015, 2016, 2018, 2019, 2020, 2021, 2022, 2023, 2024, 2025, 2026, 2027, 2028, 4001, 4022, 4023, 4024, 4025,. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
This issue is only relevant for customers importing xml files into the MagNA Pure 96 IVD instrument software for sample identification. There is a potential for sample mismatch when using the MagNA Pure 96 IVD instrument (catalog number 06541089001). Importing an order file in xml format to the MagNA Pure 96 IVD instrument software may result in a wrong sequence of samples in the graphical overvi
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Roche Diagnostics Operations, Inc.
- Distributor
- US Distribution including the states of AL, AZ, CA, CT, GA, NY, NC, NJ, OH, TX, VA and WA.
- Lot
- Serial Numbers 2001, 2002, 2003, 2004, 2006, 2011, 2012, 2013, 2014, 2015, 2016, 2018, 2019, 2020, 2021, 2022, 2023, 2024, 2025, 2026, 2027, 2028, 4001, 4022, 4023, 4024, 4025,
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAcobas HCYS, Homocysteine Enzymatic Assay, Material Number REF 06542921190, for use on cobas c 701/702, inNovember 27, 2024
- FDACreatine Kinase (CK) used on cobas c 311/501/502 and COBAS INTEGRA 400 plus-In vitro test for the quantitativeOctober 2, 2024
- FDAAlbumin Gen.2, Catalog 05166861190October 21, 2020
- FDAElecsys Anti-HAV IgM assay-for the in vitro qualitative detection of IgM antibodies to hepatitis A virusDecember 18, 2019
- FDA3M ESPE Imprint II Regular Body , Catalog Number 9379, and Paradigm Regular Body VPS, Catalog Number 5315,December 25, 2013
- FDAStreptex* kit, Ref R30950501 ZL50, Rapid latex test for Streptococcal grouping, IVD. The kit contains reagentDecember 25, 2013
Questions about this recall
Was MagNA Pure 96 System Version 1.0 Software Version 2.0 Robotic workstation for automated isolation of nucleic acids from different kinds of crude sample material (whole blood, serum, blood cells, culture cells, tissue, bacteria, fungi) by the use of the specially designed MagNA Pure reagent kits, automated filling of different kinds of PCR reaction vessels. recalled?
Yes. FDA published recall Z-0536-2014 on December 25, 2013. Reported hazard: This issue is only relevant for customers importing xml files into the MagNA Pure 96 IVD instrument software for sample identification. There is a potential for sample mismatch when using the MagNA Pure 96 IVD instrument (catalog number 06541089001). Importing an order file in xml format to the MagNA Pure 96 IVD instrument software may result in a wrong sequence of samples in the graphical overvi
What should I do if I have this product?
Voluntary: Firm initiated · Letter