Back to search
FDAFDA · Food and drugs·Recall number Z-0536-2016

ProScreen CLIA Waived Cup with 6 Drugs, Item No. PSCup-6BO-W

Published December 30, 2015 · Recall initiated: November 16, 2015

Summary

FDA records medical device Class III recall Z-0536-2016, published December 30, 2015 and initiated November 16, 2015. Recalling firm: Ameditech Inc; Status: Terminated; Reported quantity: 492 kits; Lot: 141779 142656 142963 143874 144562 151996. Class III means the product is not likely to cause adverse health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Ameditech Inc
Distributor
Nationwide Distribution.
Lot
141779 142656 142963 143874 144562 151996
Risk classification
Class III
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was ProScreen CLIA Waived Cup with 6 Drugs, Item No. PSCup-6BO-W recalled?

Yes. FDA published recall Z-0536-2016 on December 30, 2015. Reported hazard: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.

What should I do if I have this product?

Voluntary: Firm initiated · Letter