Persona The Personalized Knee System, Natural Tibia, Cemented, 5 Degree Stemmed Tivanium Sterile SIZE C, LEFT; SIZE C RIGHT This device is indicated for patients with severe knee pain and disability due to: o Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. o Collagen disorders, and/or avascular necrosis of the femoral condyle. o Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy. o Moderate valgus, varus, or flexion deformities.
Published December 26, 2012 · Recall initiated: November 14, 2012
Summary
FDA records medical device Class II recall Z-0539-2013, published December 26, 2012 and initiated November 14, 2012. Recalling firm: Zimmer, Inc.; Status: Terminated; Reported quantity: 318; Lot: Part Number 42-5320-064-01 TIBIA SIZE C, LEFT Lots: 62018336, 62018337, 62033380, 62051020, 62060739, 62068413, 62076438, 62079740, 62091360, 62117665, and 77001708, Part Number 42-5320-064-02 TIBIA SIZE C RIGHT Lot: 62018338, 62018339, 62033383, 62051021, 62060740, 62060742, 62068415, 62076439, 62079741, 62091361,and 62117666. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
All lots of the Persona Cemented Tibial Drill, 5 Degree Cemented Stemmed Natural Tibia implants, and associated tools including the Tibial Drill Guide, Broach, and Sizing Plate are being removed and replaced with enhanced products and instruments to reduce risks of tibial perforation which could result in tibial bone fracture due to the stress riser created by a perforation; tibial plate loosening
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Zimmer, Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) and internationally to Germany and Europe.
- Lot
- Part Number 42-5320-064-01 TIBIA SIZE C, LEFT Lots: 62018336, 62018337, 62033380, 62051020, 62060739, 62068413, 62076438, 62079740, 62091360, 62117665, and 77001708, Part Number 42-5320-064-02 TIBIA SIZE C RIGHT Lot: 62018338, 62018339, 62033383, 62051021, 62060740, 62060742, 62068415, 62076439, 62079741, 62091361,and 62117666
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOff-Axis Comprehensive Shoulder System Small, Augment, Off-Axis, Reamer Guide, Model/Catalog Number: 110040240July 1, 2026
- FDABrand Name: Persona Revision Product Name: Persona Revision® Trabecular Metal® Femoral Distal AugmentJuly 1, 2026
- FDAOff-Axis Alliance Glenoid, Left, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 110040444July 1, 2026
- FDAOff-Axis Comprehensive Shoulder System, Medium, Augment, Off-Axis, Reamer Guide Model/Catalog Number:July 1, 2026
- FDACoonrad/Morrey New Ulnar Revision Kit for Use with Non-Interchangeable Regular Humerals STERILE Rx, TivaniumDecember 26, 2012
- FDAPersona The Personalized Knee System, Natural Tibia, Cemented, 5 Degree Stemmed Tivanium Sterile This deviceDecember 26, 2012
Questions about this recall
Was Persona The Personalized Knee System, Natural Tibia, Cemented, 5 Degree Stemmed Tivanium Sterile SIZE C, LEFT; SIZE C RIGHT This device is indicated for patients with severe knee pain and disability due to: o Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. o Collagen disorders, and/or avascular necrosis of the femoral condyle. o Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy. o Moderate valgus, varus, or flexion deformities. recalled?
Yes. FDA published recall Z-0539-2013 on December 26, 2012. Reported hazard: All lots of the Persona Cemented Tibial Drill, 5 Degree Cemented Stemmed Natural Tibia implants, and associated tools including the Tibial Drill Guide, Broach, and Sizing Plate are being removed and replaced with enhanced products and instruments to reduce risks of tibial perforation which could result in tibial bone fracture due to the stress riser created by a perforation; tibial plate loosening
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit