Dimension(R) Chemistry Wash (RD701), Siemens Material Number (SMN) 10445052 Product Usage: Used in the Dimension(R) system with the heterogeneous module
Published January 1, 2014 · Recall initiated: May 6, 2013
Summary
FDA records medical device Class II recall Z-0561-2014, published January 1, 2014 and initiated May 6, 2013. Recalling firm: Siemens Healthcare Diagnostics, Inc.; Status: Terminated; Reported quantity: 17,719; Lot: Lot numbers RD23031, exp 2013-04-29; RD23111, exp 2013-05-07; RD23241, exp 2013-05-19; RD23311, exp 2013-06-04; and RD23461, exp 2013-06-11. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Siemens Healthcare Diagnostics received complaints of Chemistry Wash causing QC and patient sample result shifts on CTN1, LTN1, FT4, TSH, PBNP and LPBN when using Chemistry Wash lots RD23031, RD23111, RD23241, RD23311, RD23391, and RD23461. These lots shipped between 12/12/2012 and 3/22/2013. Internal investigation has confirmed shifts can be encountered in two scenarios: 1 - When an affected l
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Distributor
- US Nationwide and OUS.
- Lot
- Lot numbers RD23031, exp 2013-04-29; RD23111, exp 2013-05-07; RD23241, exp 2013-05-19; RD23311, exp 2013-06-04; and RD23461, exp 2013-06-11
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAtellica CH Revised C-Reactive Protein (RCRP); Siemens Material Number (SMN): 11537223;October 7, 2026
- FDABeckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mLAugust 26, 2026
- FDASyva EMIT 2000 Quinidine Assay REF: 4Q019UL Contents: 1x28 mL, 1x14mLAugust 26, 2026
- FDASyva Emit tox Serum Tricyclic Antidepressant Calibrator/Controls SMN 10445390 Catalog No. 7C219ULAugust 12, 2026
- FDAEdwards Lifesciences Ergonic Double Acting DeBakey Forceps with Narrow Straight Tip. Model Number ERGF01330.January 1, 2014
- FDAZiehm Vista/Vista Plus C-Arm Interventional Fluoroscopic X-Ray System. Mobile Fluoroscopic ImagingJanuary 1, 2014
Questions about this recall
Was Dimension(R) Chemistry Wash (RD701), Siemens Material Number (SMN) 10445052 Product Usage: Used in the Dimension(R) system with the heterogeneous module recalled?
Yes. FDA published recall Z-0561-2014 on January 1, 2014. Reported hazard: Siemens Healthcare Diagnostics received complaints of Chemistry Wash causing QC and patient sample result shifts on CTN1, LTN1, FT4, TSH, PBNP and LPBN when using Chemistry Wash lots RD23031, RD23111, RD23241, RD23311, RD23391, and RD23461. These lots shipped between 12/12/2012 and 3/22/2013. Internal investigation has confirmed shifts can be encountered in two scenarios: 1 - When an affected l
What should I do if I have this product?
Voluntary: Firm initiated · Letter