The Stryker Model 6390, Power-LOAD, is a power-loading cot fastener system designed to lift, lower, or steer ambulance cots into and out of the ambulance.
Published December 26, 2012 · Recall initiated: October 11, 2012
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Model 6390 Power-LOAD systems built between September 12, 2012 and September 27, 2012 may have been built with a load switch slide that does not meet specifications causing the lifting ability of the arms to be negatively impacted.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Recall details
- Manufacturer
- Stryker Medical Division of Stryker Corporation
- Distributor
- USA, Nationwide including the states of PA, NM, MI, IA, CA, CO, WA, VA, IL, WI, NY, WY, MO, TX, and CT.
- Lot
- Item 6390000000 Serial Numbers: 120940217 120940257 120940259 120940260 120940261 120940275 120940276 120940277 120940290 120940291 120940292 120940293 120940294 120940431 120940435 120940436 120940437 120940448 120940449 120940454 120940458 120940459 120940203 120940258
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States