Back to search
FDAFDA · Food and drugs·Recall number Z-0566-2021

Microsurgical Operating Room Tables;7300B (18-350-70);

Published December 23, 2020 · Recall initiated: September 11, 2019

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Unintended tilting of the Tabletop MST 7300BX and incorrect operation due to infrared operating signal errors of Tabletops MST 7300B and MST 7300BX.

What to do

Voluntary: Firm initiated · Telephone

Recall details

Manufacturer
Mizuho America, Inc.
Distributor
US: AZ, CA, FL, MO, NC, OH, PA, TX, and WI. OUS: Greece, Israel, Japan, and Ukraine.
Lot
MUC-20168-0001, MUC-20168-0002, MUC-20171-0003, MUC-20171-0004, MUC-20173-0005, MUC-20173-0006, MUC-20178-0007 & MUC-20178-0008
Risk classification
Class II
Area (FDA)
device
Country
United States