BD Vacutainer PLUS K2EDTA Tubes (Plastic). 13 x 75 mm x 2.0 mL (Pink) BD Hemogard Additive: K2EDTA (spray dried), 3.6mg Sterile IVD BD, Made in USA BD Vacutainer® tubes are used to collect, transport and process blood in a closed system. These tubes are used for testing serum, plasma or whole blood in the clinical laboratory.The BD Vacutainer® Plus K2EDTA Tubes may be used for routine immunohematology testing and blood donor screening.
Published January 6, 2016 · Recall initiated: December 3, 2015
Summary
FDA records medical device Class II recall Z-0568-2016, published January 6, 2016 and initiated December 3, 2015. Recalling firm: Becton Dickinson & Company; Status: Terminated; Reported quantity: 29,815,600 units in total; Lot: Catalog Reference #368047; Lot number 5154647. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Some of the Blood Collection Tubes in these lots were manufactured with insufficient EDTA additive. An insufficient EDTA additive may lead to erroneous platelet results, specimen recollection, or a delay of treatment.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Becton Dickinson & Company
- Distributor
- Worldwide Distribution: US (Nationwide) and the countries of: European Union, Peoples Republic of China, Japan and Republic of Korea.
- Lot
- Catalog Reference #368047; Lot number 5154647
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABD®Spinal Tray with BD®Whitacre Needle 25 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) withJune 10, 2026
- FDABD®Spinal Tray with BD®Quincke Needle 26 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) withJune 10, 2026
- FDABD®Spinal Tray with BD®Whitacre Needle 22 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) withJune 10, 2026
- FDABD®Spinal Tray with BD®Whitacre Needle 25 G x 3.5 in.; lidocaine HCL (1%), 2 mL, bupivacaine HCL (0.75%) withJune 10, 2026
- FDARenovoCath RC 120; Temporary Intravascular Occluding Catheter Model number: MMRC120-DB-1111; Rx only.January 6, 2016
- FDABD Vacutainer PLUS K2EDTA Tubes (Plastic). 13 x 75 mm x 3.0 mL (Lavender) BD Hemogard Additive: K2EDTA (sprayJanuary 6, 2016
Questions about this recall
Was BD Vacutainer PLUS K2EDTA Tubes (Plastic). 13 x 75 mm x 2.0 mL (Pink) BD Hemogard Additive: K2EDTA (spray dried), 3.6mg Sterile IVD BD, Made in USA BD Vacutainer® tubes are used to collect, transport and process blood in a closed system. These tubes are used for testing serum, plasma or whole blood in the clinical laboratory.The BD Vacutainer® Plus K2EDTA Tubes may be used for routine immunohematology testing and blood donor screening. recalled?
Yes. FDA published recall Z-0568-2016 on January 6, 2016. Reported hazard: Some of the Blood Collection Tubes in these lots were manufactured with insufficient EDTA additive. An insufficient EDTA additive may lead to erroneous platelet results, specimen recollection, or a delay of treatment.
What should I do if I have this product?
Voluntary: Firm initiated · Letter