Ziehm Solo FD. Interventional fluoroscopic x-ray system
Published December 11, 2024 · Recall initiated: August 8, 2024
Summary
FDA records medical device Class II recall Z-0574-2025, published December 11, 2024 and initiated August 8, 2024. Recalling firm: Orthoscan, Inc.; Status: Ongoing; Reported quantity: 25; Lot: UDI: EZIEZIEHMSOLOFD1 Serial Numbers: 54301, 54772, 54773, 54774, 54792, 54793, 54794, 54795, 54796, 54797, 54798, 54851, 54852, 54855, 55020, 55021, 55114, 55115, 55116, 55117, 55118, 55119, 55120, 55121, 55212. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Ziehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with respect to wired hand switch may fail when being exposed to mechanical stress such as shocks (e.g. dropping) or impacts can irreversibly damage the switching element in the hand switch. A defective switching element may then potentially initiate unintended radiation of the connected mobile X-ray unit.
What to do
FDA Mandated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Orthoscan, Inc.
- Distributor
- U.S. (including Puerto Rico).
- Lot
- UDI: EZIEZIEHMSOLOFD1 Serial Numbers: 54301, 54772, 54773, 54774, 54792, 54793, 54794, 54795, 54796, 54797, 54798, 54851, 54852, 54855, 55020, 55021, 55114, 55115, 55116, 55117, 55118, 55119, 55120, 55121, 55212.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAZiehm Solo. Digital Mobile C-armDecember 11, 2024
- FDAZiehm Vision system, x-ray, tomography, computed.December 11, 2024
- FDAZiehm Vision R. Iterventional fluoroscopic x-ray systemDecember 11, 2024
- FDAZiehm Vision RFD 3D. Interventinoal fluoroscopic x-ray systemDecember 11, 2024
- FDADiode Laser Hair Removal model: WLA-01December 11, 2024
- FDAZiehm Vision FD. Interventional fluoroscopic x-ray systemDecember 11, 2024
Questions about this recall
Was Ziehm Solo FD. Interventional fluoroscopic x-ray system recalled?
Yes. FDA published recall Z-0574-2025 on December 11, 2024. Reported hazard: Ziehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with respect to wired hand switch may fail when being exposed to mechanical stress such as shocks (e.g. dropping) or impacts can irreversibly damage the switching element in the hand switch. A defective switching element may then potentially initiate unintended radiation of the connected mobile X-ray unit.
What should I do if I have this product?
FDA Mandated · Letter