Laser Fiber Assy-Single Use-LFS-272-RT-0.22NA-SMA905- Slim Extension Sleeve Silver-Sterile ETO- Private label: SU-200-RT SmartScope Packaging. The Laser peripherals bare laser fibers, ENT fibers and Endoprobes are intended for use in laser surgical procedure for cutting, coagulating, or vaporizing in any soft tissue application for which compatible Nd: YAG, Ho: YAG, KTP and Diode lasers have been cleared.
Published December 19, 2018 · Recall initiated: October 17, 2018
Summary
FDA records medical device Class II recall Z-0577-2019, published December 19, 2018 and initiated October 17, 2018. Recalling firm: Leoni Fiber Optics GmbH; Status: Terminated; Reported quantity: N/A; Lot: 096874 00816901020265. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A percentage of the products shipped in Smartscope single-box packaging may exhibit a broken package seal, thus having compromised sterility. There have been two reports of compromised sterility; however, all products packaged in Smartscope single-box packaging are at risk of a compromised product seal. Broken seal pouches compromise sterility and could increase the risk of infection if used.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Leoni Fiber Optics GmbH
- Distributor
- US Distribution to the state of: Ohio.
- Lot
- 096874 00816901020265
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Germany
Related recalls
- FDALaser Fiber Assy-Single Use-LFS-200-0.22NA-SMA905-Extension Sleeve Silver-Sterile ETO-ForTec FiberDecember 19, 2018
- FDALaser Fiber Assy-Single Use-LFS-940-RT-0.22NA-SMA905-Extension Sleeve Silver-Sterile ETO-Private label:December 19, 2018
- FDACIVCO Sales Demo Kit, REF DEMOGPGUIDE, for use by CIVCO and customer sales representatives.December 19, 2018
- FDASonoSite FUJIFILM Needle Guide Bracket, REF number P06013-08, Reusable non-sterile bracket for use withDecember 19, 2018
- FDASalmon fillets packaged inside Styrofoam boxes fully surrounded and covered in ice, which were stored inside aDecember 19, 2018
- FDAPROCARE B20 Patient Monitor, V1; and B20 Patient Monitor, V2. The PROCARE Monitor B20 is a portableDecember 19, 2018
Questions about this recall
Was Laser Fiber Assy-Single Use-LFS-272-RT-0.22NA-SMA905- Slim Extension Sleeve Silver-Sterile ETO- Private label: SU-200-RT SmartScope Packaging. The Laser peripherals bare laser fibers, ENT fibers and Endoprobes are intended for use in laser surgical procedure for cutting, coagulating, or vaporizing in any soft tissue application for which compatible Nd: YAG, Ho: YAG, KTP and Diode lasers have been cleared. recalled?
Yes. FDA published recall Z-0577-2019 on December 19, 2018. Reported hazard: A percentage of the products shipped in Smartscope single-box packaging may exhibit a broken package seal, thus having compromised sterility. There have been two reports of compromised sterility; however, all products packaged in Smartscope single-box packaging are at risk of a compromised product seal. Broken seal pouches compromise sterility and could increase the risk of infection if used.
What should I do if I have this product?
Voluntary: Firm initiated · Letter