DIST LAT FEM LOCK PLT, RT

Published January 8, 2014 · Recall initiated: November 5, 2013

Summary

FDA records medical device Class II recall Z-0588-2014, published January 8, 2014 and initiated November 5, 2013. Recalling firm: Zimmer, Inc.; Status: Terminated; Reported quantity: 42, 065 units total; Lot: Item Numbers: 47-2357-101-14 47-2357-101-18 47-2357-102-14 47-2357-102-18 All lots with an expiration date prior to July 24, 2023. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Based on a review of historic packaging validations, it was determined that a subset of product packaged at a Zimmer facility required testing to ensure that previously packaged product from this facility met the established requirements for sterile barrier integrity. This testing determined that a packaging tray configuration which consists of an inner and outer sterile tray, resulted in failure

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Zimmer, Inc.
Distributor
Worldwide Distribution.
Lot
Item Numbers: 47-2357-101-14 47-2357-101-18 47-2357-102-14 47-2357-102-18 All lots with an expiration date prior to July 24, 2023
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was DIST LAT FEM LOCK PLT, RT recalled?

Yes. FDA published recall Z-0588-2014 on January 8, 2014. Reported hazard: Based on a review of historic packaging validations, it was determined that a subset of product packaged at a Zimmer facility required testing to ensure that previously packaged product from this facility met the established requirements for sterile barrier integrity. This testing determined that a packaging tray configuration which consists of an inner and outer sterile tray, resulted in failure

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail