Back to search
FDAFDA · Food and drugs·Recall number Z-0590-2020

CrossCath Support Catheter, RPN CXC3.4-2.2-18-150-P-NS-0, Order Number G51577 For diagnostic and interventional vascular procedures.

Published December 18, 2019 · Recall initiated: November 8, 2019

Summary

FDA records medical device Class I recall Z-0590-2020, published December 18, 2019 and initiated November 8, 2019. Recalling firm: Cook Inc.; Status: Terminated; Reported quantity: 58; Lot: Lot Numbers 9950878, 9953506 UDI: (01)00827002515772(17)221508(10)9950878 UDI: (01)00827002515772(17)221608(10)9953506. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A manufacturing error may cause the radiopaque marker bands to be too loose on certain CXC3.0 CrossCath Support Catheters (compatible with 0.014 wire guides) and too tight on certain CXC3.4 CrossCath Support Catheters (compatible with 0.018 wire guides).

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Cook Inc.
Distributor
Distribution to US states of AL, CA, CT, FL, GA, IL, MA, MD, MI, MO, NC, NJ, NY, OH, OK, PA, SC, TN, TX, and VA, and Brazil.
Lot
Lot Numbers 9950878, 9953506 UDI: (01)00827002515772(17)221508(10)9950878 UDI: (01)00827002515772(17)221608(10)9953506
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was CrossCath Support Catheter, RPN CXC3.4-2.2-18-150-P-NS-0, Order Number G51577 For diagnostic and interventional vascular procedures. recalled?

Yes. FDA published recall Z-0590-2020 on December 18, 2019. Reported hazard: A manufacturing error may cause the radiopaque marker bands to be too loose on certain CXC3.0 CrossCath Support Catheters (compatible with 0.014 wire guides) and too tight on certain CXC3.4 CrossCath Support Catheters (compatible with 0.018 wire guides).

What should I do if I have this product?

Voluntary: Firm initiated · Letter