CrossCath Support Catheter, RPN CXC3.4-2.2-18-150-P-NS-0, Order Number G51577 For diagnostic and interventional vascular procedures.
Published December 18, 2019 · Recall initiated: November 8, 2019
Summary
FDA records medical device Class I recall Z-0590-2020, published December 18, 2019 and initiated November 8, 2019. Recalling firm: Cook Inc.; Status: Terminated; Reported quantity: 58; Lot: Lot Numbers 9950878, 9953506 UDI: (01)00827002515772(17)221508(10)9950878 UDI: (01)00827002515772(17)221608(10)9953506. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A manufacturing error may cause the radiopaque marker bands to be too loose on certain CXC3.0 CrossCath Support Catheters (compatible with 0.014 wire guides) and too tight on certain CXC3.4 CrossCath Support Catheters (compatible with 0.018 wire guides).
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Cook Inc.
- Distributor
- Distribution to US states of AL, CA, CT, FL, GA, IL, MA, MD, MI, MO, NC, NJ, NY, OH, OK, PA, SC, TN, TX, and VA, and Brazil.
- Lot
- Lot Numbers 9950878, 9953506 UDI: (01)00827002515772(17)221508(10)9950878 UDI: (01)00827002515772(17)221608(10)9953506
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAFlexor Check-Flo Introducer - Raabe Modification, Introducer Set, RPN KCFW-5.0-38-90-RB-RAABE, GPN G13069 TheNovember 10, 2021
- FDAFlexor Check-Flo Introducer, RPN KCFW-8.0-35-55-RB-HFANL0-HC, Order Number G56233 Flexor Introducers andDecember 30, 2020
- FDAFlexor Check-Flo Introducer, RPN KCFW-6.0-35-55-RB-HFANL0-HC, Order Number G56231 Flexor Introducers andDecember 30, 2020
- FDAFlexor Check-Flo Introducer, RPN KCFW-6.0-35-55-RB-HFANL1-HC, Order Number G56236 Flexor Introducers andDecember 30, 2020
- FDATailor Cut Produce Inc., Pineapple Chunks, 2 - 1 gallon tubs per case.December 18, 2019
- FDAC7000 - CUSA® Clarity Ultrasonic Tissue Ablation SystemDecember 18, 2019
Questions about this recall
Was CrossCath Support Catheter, RPN CXC3.4-2.2-18-150-P-NS-0, Order Number G51577 For diagnostic and interventional vascular procedures. recalled?
Yes. FDA published recall Z-0590-2020 on December 18, 2019. Reported hazard: A manufacturing error may cause the radiopaque marker bands to be too loose on certain CXC3.0 CrossCath Support Catheters (compatible with 0.014 wire guides) and too tight on certain CXC3.4 CrossCath Support Catheters (compatible with 0.018 wire guides).
What should I do if I have this product?
Voluntary: Firm initiated · Letter