FDAFDA · Food and drugs·Recall number Z-0591-2024
Omnipod 5 App (on compatible Android smartphones), Software Versions 1.1 - 1.2.3 (current). US: PT-000599- Omnipod 5 App Software Package (provided through Google Play), EU: PT-001323- Omnipod 5 App Software Package (provided on locked-down Controllers (PT-001194 and PT-001193). Used to control an insulin pump.
Published January 10, 2024 · Recall initiated: November 30, 2023
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The bolus calculator is not recording the decimal point if it is the first value entered when changing a bolus dose. This may lead to an over-delivery of insulin to the user if the user does not recognize the error on the bolus calculator screen or the confirmation screen prior to starting the bolus.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Insulet Corporation
- Distributor
- Worldwide distribution - US Nationwide and the countries of Germany and UK.
- Lot
- UDI/Model/NDC Number: " Omnipod 5 App Software UDI: 1038508112030 " EU locked-down Controller UDI: 10385083000190 and 10385083000206
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States