Back to search
FDAFDA · Food and drugs·Recall number Z-0591-2024

Omnipod 5 App (on compatible Android smartphones), Software Versions 1.1 - 1.2.3 (current). US: PT-000599- Omnipod 5 App Software Package (provided through Google Play), EU: PT-001323- Omnipod 5 App Software Package (provided on locked-down Controllers (PT-001194 and PT-001193). Used to control an insulin pump.

Published January 10, 2024 · Recall initiated: November 30, 2023

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The bolus calculator is not recording the decimal point if it is the first value entered when changing a bolus dose. This may lead to an over-delivery of insulin to the user if the user does not recognize the error on the bolus calculator screen or the confirmation screen prior to starting the bolus.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Insulet Corporation
Distributor
Worldwide distribution - US Nationwide and the countries of Germany and UK.
Lot
UDI/Model/NDC Number: " Omnipod 5 App Software UDI: 1038508112030 " EU locked-down Controller UDI: 10385083000190 and 10385083000206
Risk classification
Class I
Area (FDA)
device
Country
United States